NUK - logo
E-viri
  • Efficacy and Safety of Tran...
    Xie, Xiaochuan; Shi, Xiaohan; Xun, Xiaoshuang; Rao, Li

    Annals of Thoracic and Cardiovascular Surgery, 01/2016, Letnik: 22, Številka: 4
    Journal Article

    Purpose: To elucidate the performance of transcatheter aortic valve implantation (TAVI) in bicuspid aortic valve (BAV) patients through a systematic review and meta-analysis.Methods: A systematic literature review was performed by searching eligible articles in PubMed, Medline, EMBASE, Google Scholar and CNKI. Meta-analysis of included case-control/cohort studies was further conducted. Relative risks (RRs) and the corresponding 95% confidence intervals (CIs) were used to compare clinical outcomes of BAV patients and non-BAV patients.Results: A total of 17 articles including eight case reports, four case series and five case-control/cohort studies with 166 BAV patients were analyzed. Device success rate achieved for TAVI in this cohort of BAV patients was 95.2%. The 30-day mortality rate was 8.4%, and the medium-term (range from 6 months to 2 years) mortality rate reported was 17.9%. Overall, the performance of TAVI in BAV patients was comparable to that in non-BAV patients, as reported by the included case-control/cohort studies (30-day mortality rate: RR = 1.05, 95%CI 0.57–1.95, p = 0.87; Device success rate: RR = 1.00, 95%CI 0.95–1.05, p = 0.94; Incidence of moderate to severe paravalvular regurgitation: RR = 1.25, 95%CI 0.85–1.84, p = 0.25).Conclusion: The present study suggested that TAVI may be a feasible and safe treatment modality for BAV patients.