ALL libraries (COBIB.SI union bibliographic/catalogue database)
PDF
  • High-sensitivity, simple frequency-resolved-optical-gating device
    Jones, Travis ...
    We introduce a practical self-referenced device for the complete temporal intensity-and-phase measurement of fewfemtojoule-level fs and ps pulses based on the simple version of ... frequency-resolved-optical gating, called GRENOUILLE. We replace the usual crossed-beam line-focus geometry with a pointfocus near-collinear-beam geometry and a traditional delay line that allows a longer path length in the nonlinear medium, yielding greater sensitivity. As an initial demonstration of the device, we measured moderately complex pulses at energies as weak as 24 fJ with high signal-to-noise ratio. We confirmed the results measured using our technique to that of a standard GRENOUILLE and a high-resolution spectrometer.
    Source: IEEE journal of quantum electronics. - ISSN 0018-9197 (Vol. 56, iss. 3, 2020, str. 1-6)
    Type of material - article, component part
    Publish date - 2020
    Language - english
    COBISS.SI-ID - 16641283