ALL libraries (COBIB.SI union bibliographic/catalogue database)
  • Gamma-ray spectroscopy at MHz counting rates with a compact LaBr3 detector and silicon photomultipliers for fusion plasma applications
    Nocente, Massimo ...
    Gamma-ray spectroscopy measurements at MHz counting rates have been carried out, for the first time, with a compact spectrometer based on a LaBr3 scintillator and silicon photomultipliers. The ... instrument, which is also insensitive to magnetic fields, has been developed in view of the upgrade of the gamma-ray camera diagnostic for a particle measurements in deuterium-tritium plasmas of the Joint European Torus. Spectra were measured up to 2.9 MHz with a projected energy resolution of 3%-4% in the 3-5 MeV range, of interest for fast ion physics studies in fusion plasmas. The results reported here pave the way to first time measurements of the confined a particle profile in high power plasmas of the next deuterium-tritium campaign at the Joint European Torus
    Source: Proceedings (Iss. 11, Vol. 87, 2016, Article 11E714)
    Type of material - conference contribution
    Publish date - 2016
    Language - english
    COBISS.SI-ID - 30724135
    DOI

source: Proceedings (Iss. 11, Vol. 87, 2016, Article 11E714)