  • Analysis of loading profile effect on testing machine calibration results [Elektronski vir]
    Hiti, Miha
    The paper presents an evaluation of the loading profile effect on the testing machine calibration. First, a 20 kN force transducer was calibrated in a force calibration machine. This transducer was ... then used for a number of calibration runs in a 600 kN material testing machine where the time step interval of the same loading profile was varied from 120 s to 0 s. Results show good agreement for most of the range but highlight an increasing error due to synchronization and filtering issues in measured signals when approaching very short step intervals and continuous loading.
    Source: Measurement [Elektronski vir]. Sensors. - ISSN 2665-9174 (Vol. 18, Dec. 2021, str. 1-5)
    Type of material - e-article ; adult, serious
    Publish date - 2021
    Language - english