  • Fluence profiling at JSI TRIGA reactor irradiation facility
    Sola, Valentina ...
    We present an analysis of the fluence profile at the JSI TRIGA neutron reactor facility in Ljubljana. For the study, multi-pad Low-Gain Avalanche Diodes (LGADs) are used. The deactivation of acceptor ... doping in the gain layer implant due to the irradiation, typical of LGAD devices, is exploited to map the fluence profile inside the irradiation channels. The amount of active doping of the LGAD gain layer is extracted via capacitance–voltage measurements for each pad before and after irradiation to a fluence of 1.5 × 1015 neq∕cm2 , where 𝑛𝑒𝑞 stands for 1 MeV equivalent neutron count, providing a precise and prompt measurement of the fluence distribution over the LGAD sensor. Experimental results are compared to neutron fluence expectations calculated with Monte Carlo techniques.
    Vrsta gradiva - članek, sestavni del
    Leto - 2024
    Jezik - angleški
    COBISS.SI-ID - 192146947
    DOI