  • Ion trap mass spectrometry as detector for capillary electrochromatography of peptides : possibilities and limitations
    Gaspari, Marco ...
    Capillary electrochromatography has been coupled to an ion trap mass spectrometer via a sheath flow electrospray interface. Peptide analysis was perfomed on 25 cm columns packed with ▫$C_18$▫-bonded ... silica particles, and very fast separations were obtained. Full mass spectroscopy and tandem mass spectrometry (MS/MS) detection were evaluated in terms of sensitivity and scanning speed. The separation of seven selected peptides and their detection by MS/MS in selected reaction monitoring mode was demonstrated. The performance of this sheath flow interface also was compared with the previously reported sheathless interface. Despite the loss in sensitivity (20-40-fold), the sheath flow interface was shown to be superior in terms of ruggedness, and allowed the use of higher electric fields to achieve faster analysis times.
    Vir: The Journal of microcolumn separations. - ISSN 1040-7685 (Vol. 13, is. 6, 2001, str. 243-249)
    Vrsta gradiva - članek, sestavni del
    Leto - 2001
    Jezik - angleški
    COBISS.SI-ID - 23818245