  • Pulse wave analysis to assess systemic blood flow during mechanical biventricular support
    Scolletta, S ...
    Measurement of systemic blood flow is of crucial importance in patients on mechanical circulatory support (MCS). We reported the case of a 65-year-old female patient in severe cardiogenic shock ... undergoing left (Jarvik 2000 axial flow pump) and right (Levitronix-Centrimag centrifugal pump) ventricular assist device implant. Evaluation of blood flow was obtained by ultrasonic flowmetry, continuous thermodilution technique, and pressure recording analytical method (PRAM). This pulse contour system allows beat-by-beat systemic blood flow assessment from the analysis of radial artery pressure waveform. At a Jarvik pump speed < or = 10000 rotations per minutes (rpm), thermodilution and PRAM showed similar blood flow values. At a Jarvik pump speed > or = 11000 rpm, the aortic valve did not open and PRAM did not provideblood flow values due to nonpulsatile blood flow. The present paper describes the first experience with PRAM in a single patient on MCS. Further studies are required to assess the validity of PRAM as an additional monitoring system in the setting of ventricular assist device support.
    Vir: Perfusion. - ISSN 0267-6591 (Letn. 22, št. 1, 2007, str. 63-66)