VSE knjižnice (vzajemna bibliografsko-kataložna baza podatkov COBIB.SI)
PDF
  • A novel method for the determination of field output factors and output correction factors for small static fields for six diodes and a microdiamond detector in megavoltage photon beams
    Casar, Božidar ...
    The goal of this work is to provide a large and consistent set of data for detector specific output correction factors, k fclin, fref / Qclin, Qref, for small static fields for seven solid state ... detectors and to determine field output factors, [omega]fclin, fref / Qclin, Qref, using EBT3 radiochromic films and W1 plastic scintillator as reference detectors on two different linear accelerators and four megavoltage photon beams. Consistent measurement conditions and recommendations given in the International Code of PracticeTRS-483 for small field dosimetry were followed throughout the study ... [etc].
    Vir: Medical physics. - ISSN 0094-2405 (Vol. 46, no. 2, 2019, str. 944-963)
    Vrsta gradiva - članek, sestavni del
    Leto - 2019
    Jezik - angleški
    COBISS.SI-ID - 3097211
    DOI

vir: Medical physics. - ISSN 0094-2405 (Vol. 46, no. 2, 2019, str. 944-963)