VSE knjižnice (vzajemna bibliografsko-kataložna baza podatkov COBIB.SI)
  • In-situ plated antimony film electrode for adsorptive cathodic stripping voltammetric measurement of trace nickel
    Sopha, Hanna Ingrid ...
    Antimony film electrode (SbFE) prepared in-situ is presented for the first time in combination with adsorptive cathodic stripping regime for measuring trace levels of Ni(II). Stabilization of Sb(III) ... in mildly alkaline solution (pH 9.0) was achieved through the addition of potassium sodium tartrate in themeasurement solution, and as a complexing agent for Ni(II) dimethylglyoximewas selected. The SbFE exhibited excellent electroanalytical performance in the presence of dissolved oxygen with a calculated LoD of 0.11 Žg LŽ1 Ni(II) in combination with 60 s accumulation time and favorable linear response in the examined concentration range of 2-20 Žg LŽ1 (R2=0.998) associated with only 20 s accumulation. The electrode revealed superior performance over its ex-situ prepared analogue and, furthermore, exhibited auspicious electroanalytical performance also under constant current chronopotentiometric conditions
    Vir: Electrochemistry communications. - ISSN 1388-2481 (Vol. 20, 2012, str. 23-25)
    Vrsta gradiva - članek, sestavni del
    Leto - 2012
    Jezik - angleški
    COBISS.SI-ID - 4942618