  • Fracture properties of titanium bone implants corroded by simulated body fluid = Lomne lastnosti titanovih kostnih implantantov korodiranih s simulirano telesno tekočino
    Marčič, Simon ; Praunseis, Anja
    Titanium has been widely used for medical implants and restorations due to its excellent osseointegration, corrosion resistance, biocompatibility in biological fluids, and high resistance/weight ... ratio. The aim of this paper is to determine the fracture properties of two bone implants of a hip prosthesis made from commercially pure titanium and Ti-6Al-4V titanium alloy. Both bone implants were before fracture mechanics testing corroded by simulated body fluid. The presence of different microstructures along the pre-crack fatigue front has significant effects on the critical crack tip opening displacement (CTOD). The CTOD values were calculated in accordance with standard ASTM E1290-08e1. This value can be the relevant parameter for the safe servicing of hip prostheses.
    Vir: Journal of energy technology. - ISSN 1855-5748 (Vol. 9, iss. 2, aug. 2016, str. 55-62)