  • Evaluation of silicona microstrip detectors as X-ray sensors in digital mammography = Ovrednotenje silicijevih mikropasovnih detektorjev kot rentgenskih senzorjev za digitalno mamografijo
    Mali, Tadej ...
    Background. Position sensitive silicon microstrip detectors are used as sensors for X rays in a digital imaging system. Silicon detectors were used inan edgeon geometry, yielding high X ray detection ... efficiency. Material and methods. A small detector system was assembled and tested. Images of a standard, 5 cm thick phantom were made and evaluated. It is demonstrated, thatthe use of silicon detectors in mammography could significantly contributeto a reduction of dose. All images were made with skin entrance doses lower than 1 mGy. Results and conclusion. Microcalcifications with a diameter of 350microm could still be detected with skin entrance doses of about 0.25 mGy. It was demonstrated that a 5 Ip/mm pattern can be detected. Image processing should further improve the image quality.
    Vir: Radiology and oncology. - ISSN 1318-2099 (Vol. 33, no. 3, 1999, str. 237-244)