  • Photoplethysmographic heart rate and blood pressure estimation in dynamic condition
    Zazula, Damjan ...
    This paper describes a novel, unobtrusive photoplethysmographic device and experiments conducted in a dynamic condition. The device complements a photoplethysmograph with a dynamometer, an ... accelerometer, and a thermometer. We built it in the door handle of a refrigerator and made tests during the door opening and closing. Short-term PPG signals were obtained across all fingers pulling the handle. In parallel, we measured the standard ECG lead II as a reference for the detected heartbeat verification. The blood pressure of each participant was registered before each experimental trial by a standard sphygmomanometer. Based on the PPG signals, we ran our computer algorithms for detecting the heartbeats and estimating the mean blood pressure. Satisfactory results were achieved for heartbeats, with 97.62% sensitivity and 90.31% precision. However, blood pressure estimation appeared to be not very efficient - neither linear nor exponential modeling could lower the estimated mean absolute error in comparison to referential mean blood pressures below 6 mmHg
    Vrsta gradiva - članek, sestavni del