Narodna in univerzitetna knjižnica, Ljubljana (NUK)
Naročanje gradiva za izposojo na dom
Naročanje gradiva za izposojo v čitalnice
Naročanje kopij člankov
Urnik dostave gradiva z oznako DS v signaturi
  • Material failure of an AISI 316L stainless steel hip prosthesis = Napake materiala v kolčni protezi iz nerjavnega jekla AISI 316L
    Godec, Matjaž, 1961-
    One of the major successes of modern surgery is the total replacement of joints by timplantation of prosthetic components by which the pain is relieved and deformity is corrected. The aseptic ... loosening of a prosthetic-joint component is the most common reason for joint-revision surgery. Furthemore, it is thought that wear particles are one of the major contributors to the development and perpetuation of aseptic loosening. It was ascertained that besides wear particles originated from polyethylene acetabular cup, the main reason of loosening hip prosthesis was due to the material failure. Based on stem material examination by different techniques it was found out that improper microstructure caused additional abrasion by hard particles and aseptic loosening took place.
    Vir: Materiali in tehnologije = Materials and technology. - ISSN 1580-2949 (Letn. 45, št. 2, mar.-apr. 2011, str. 85-90)
    Vrsta gradiva - članek, sestavni del
    Leto - 2011
    Jezik - angleški
    COBISS.SI-ID - 855210

vir: Materiali in tehnologije = Materials and technology. - ISSN 1580-2949 (Letn. 45, št. 2, mar.-apr. 2011, str. 85-90)