    Biegała, Michał; Jakubowska, Teresa

    Radiation protection dosimetry, 07/2020, Volume: 189, Issue: 1
    Journal Article

    Abstract This paper aims to determine the levels of exposure to neutron and photon radiation among the personnel engaged in cyclotron operation and the personnel engaged in the production of radiopharmaceuticals, with the use of the environmental radiation monitoring system (RMS) installed in the positron emission tomography laboratory. The annual exposures of employees operating the cyclotron measured with the use of the RMS system are: 1.39 ± 0.16 mSv in case of photon radiation and 2.61 ± 0.14 mSv in case of neutron radiation. In the case of employees in the radiopharmaceuticals’ production zone, the annual exposures measured by means of the RMS system are 0.15 ± 0.03 mSv in case of photon radiation and 0.11 ± 0.01 mSv in case of neutron radiation. The exposure levels among the personnel engaged in cyclotron operation and the personnel engaged in the production of radiopharmaceuticals are below the permissible radiation dose limits.