    Muff, Guillaume; Dufour, Stéphane; Meyer, Alain; Severac, François; Favret, Fabrice; Geny, Bernard; Lecocq, Jehan; Isner-Horobeti, Marie-Eve

    Journal of Physical Therapy Science, 2016, Volume: 28, Issue: 9
    Journal Article

    Purpose To compare measurements of knee extensor and flexor muscle strength performed using a hand-held dynamometer and an isokinetic dynamometer in apparently healthy subjects. Subjects and Methods Thirty adult volunteers underwent knee muscle strength evaluation using an isokinetic or a hand-held dynamometer. Results Strong positive correlations were found between the 2 methods, with correlation coefficients r ranging from 0.72 (95% confidence interval CI, 0.48−0.86) to 0.87 (95% CI, 0.75−0.94), depending on the muscle group and the isokinetic evaluation mode. The reproducibility of the hand-held dynamometer findings was good, judged by a coefficient of variation of 3.2–4.2%. However, the correlation between the 2 methods for the assessment of flexor/extensor ratios ranged from −0.04 to 0.46. Conclusion Knee extensor and flexor muscle strength recorded with a hand-held dynamometer is reproducible and significantly correlated with the isokinetic values, indicating that this method may in some cases be a useful replacement for isokinetic strength measurement. However, for strength ratio assessment, and when judged against the isokinetic standard, a hand-held dynamometer is not a valid option.