The human hand is the most frequently used body part in activities of daily living. With its complex anatomical structure and the small size compared to the body, assessing the functional capability ...is highly challenging. The aim of this review was to provide a systematic overview on currently available 3D motion analysis based on skin markers for the assessment of hand function during activities of daily living. It is focused on methodology rather than results. A systematic review according to the PRISMA guidelines was performed. The systematic search yielded 1349 discrete articles. Of 147 articles included on basis of title, 123 were excluded after abstract review, and 24 were included in the full-text analysis with 13 key articles. There is still limited knowledge about hand and finger kinematics during activities of daily living. A standardization of the task is required in order to overcome the nonrepetitive nature and high variability of upper limb motion and ensure repeatability of task performance. To yield a progress in the analysis of human hand movements, an assessment of human kinematics including fingers, wrist, and thumb and an identification of relevant parameters that characterize a healthy motion pattern during functional tasks are needed.
Measuring joint kinematics is a key requirement for a plethora of biomechanical research and applications. While x-ray based systems avoid the soft-tissue artefacts arising in skin-based measurement ...systems, extracting the object’s pose (translation and rotation) from the x-ray images is a time-consuming and expensive task. Based on about 106’000 annotated images of knee implants, collected over the last decade with our moving fluoroscope during activities of daily living, we trained a deep-learning model to automatically estimate the 6D poses for the femoral and tibial implant components. By pretraining a single stage of our architecture using renderings of the implant geometries, our approach offers personalised predictions of the implant poses, even for unseen subjects. Our approach predicted the pose of both implant components better than about 0.75 mm (in-plane translation), 25 mm (out-of-plane translation), and 2° (all Euler-angle rotations) over 50% of the test samples. When evaluating over 90% of test samples, which included heavy occlusions and low contrast images, translation performance was better than 1.5 mm (in-plane) and 30 mm (out-of-plane), while rotations were predicted better than 3−4°. Importantly, this approach now allows for pose estimation in a fully automated manner.
Purpose
Unicompartmental knee arthroplasty (UKA), resulting in similar kinematics to native knees, is functionally superior to total knee arthroplasty (TKA). However, ACL deficiency is generally ...considered to be a contraindication. The main purpose of this study was to investigate if UKA in ACL-deficient knees would result in similar kinematics to conventional UKA with an intact ACL.
Methods
Ten conventional UKA patients were compared to eight ACL-deficient patients with a reduced tibial slope to compensate for instability, resulting from the deficient ACL. Knee kinematics was evaluated with a moving fluoroscope, tracking the knee joint during daily activities. In a standing position (baseline), posterior shift of the femur was observed for ACL-deficient UKA patients, compared to conventional UKA patients.
Results
A significant posterior femoral shift in the ACL-deficient group was observed during the first 25% (near extension) of deep knee bend, while there was no difference in kinematic waveforms for all other activities. No significant range of motion differences across different activities between the two UKA groups were detected, except for an increase of medial AP translation in the ACL-deficient group, during deep knee bend and stair descent.
Conclusion
Despite the posterior femoral shift due to ACL deficiency, both UKA groups showed similar kinematic waveforms, indicating that posterior tibial slope reduction can partially compensate for ACL function. This supported our hypothesis that fixed bearing UKA can be a viable treatment option for selected ACL-deficient patients, allowing patient-specific kinematics. While anteroposterior laxity can be compensated, rotational stability was a prerequisite for this approach.
Level of evidence
III.
•New algorithm was able to register with an accuracy of <1 mm (in-plane) and <2°.•The algorithm could not register all images successfully.•Semi-automatic or manual registration is still required to ...obtain good results.
The accurate quantification of in-vivo tibio-femoral kinematics is essential for understanding joint functionality, but determination of the 3D pose of bones from 2D single-plane fluoroscopic images remains challenging. We aimed to evaluate the accuracy, reliability and repeatability of an intensity-based 2D/3D registration algorithm.
The accuracy was evaluated using fluoroscopic images of 2 radiopaque bones in 18 different poses, compared against a gold-standard fiducial calibration device. In addition, 3 natural femora and 3 natural tibiae were used to examine registration reliability and repeatability.
Both manual fitting and intensity-based registration exhibited a mean absolute error of <1 mm in-plane. Overall, intensity-based registration of the femoral bone model revealed significantly higher translational and rotational errors than manual fitting, while no statistical differences (except for y-axis translation) were found for the tibial bone model. The repeatability of 108 intensity-based registrations showed mean in-plane standard deviations of 0.23–0.56 mm, but out-of-plane position repeatability was lower (mean SD: femur 7.98 mm, tibia 6.96 mm). SDs for rotations averaged 0.77–2.52°.
While the algorithm registered some images extremely well, other images clearly required manual intervention. When the algorithm registered the bones repeatably, it was also accurate, suggesting an approach that includes manual intervention could become practical for efficient and accurate registration.
The measurement of finger and wrist range of motion (ROM) is of great importance to clinicians when assessing functional outcomes of therapeutic interventions and surgical procedures. The purpose of ...the study was to assess the repeatability of ROM measurements of the hand joints with manual goniometer and 3D motion capture system and to calculate the minimal detectable difference for both methods.
Active finger and wrist joints ROM of 20 healthy volunteers were assessed using a manual goniometer and 3D motion capture system. Minimal detectable difference (MDD) and standard error of measurement (SEM) were calculated for both measurement systems and compared within the same task. Maximal ROM of all joints was registered twice on two different days to evaluate the test-retest repeatability. The intraclass correlation coefficients (ICC) was calculated and examined to determine if reliability ≥ 0.70 existed.
MDD for the 3D motion capture was between 5 and 12° except for the metacarpophalangeal joint (MCP) 1, interphalangeal joint (IP), and MCP5. SEM values lay between 2 and 4° for all joints except for the MCP5, IP, and MCP1. For the goniometric measurements, MDD and SEM were between 12-30° and 4-11°, respectively. The reliability criterion (ICC > 0.7) was achieved for the ROM measurement with the 3D motion capture system for 94% of the joints and in only 65% of the joints with the manual goniometer.
Joint ROM assessed with 3D motion analysis showed higher test-retest agreement demonstrating overall better repeatability for this method. Because of the smaller measurement error, the 3D motion capture system has a smaller MDD. Only individual test-rest differences bigger than the MDD can be considered as real changes, and therefore, in an experimental situation, the use of a more precise measurement method can greatly reduce the number of subjects needed for a statistical significance. Goniometer measurements of some joints should be carefully interpreted, due to a low repeatability and reliability.
This study is approved by the Ethical Committee Zurich ( Kek-ZH-Nr: 2015-0395 ).
Videofluoroscopy has been shown to provide essential information in the evaluation of the functionality of total knee arthroplasties. However, due to the limitation in the field of view, most systems ...can only assess knee kinematics during highly restricted movements. To avoid the limitations of a static image intensifier, a moving fluoroscope has been presented as a standalone system that allows tracking of the knee during multiple complete cycles of level- and downhill-walking, as well as stair descent, in combination with the synchronous assessment of ground reaction forces and whole body skin marker measurements. Here, we assess the ability of the system to keep the knee in the field of view of the image intensifier. By measuring ten total knee arthroplasty subjects, we demonstrate that it is possible to maintain the knee to within 1.8 ± 1.4 cm vertically and 4.0 ± 2.6 cm horizontally of the centre of the intensifier throughout full cycles of activities of daily living. Since control of the system is based on real-time feedback of a wire sensor, the system is not dependent on repeatable gait patterns, but is rather able to capture pathological motion patterns with low inter-trial repeatability.
The success of kinematic analysis that relies on inertial measurement units (IMUs) heavily depends on the performance of the underlying algorithms. Quantifying the level of uncertainty associated ...with the models and approximations implemented within these algorithms, without the complication of soft-tissue artefact, is therefore critical. To this end, this study aimed to assess the rotational errors associated with controlled movements. Here, data of six total knee arthroplasty patients from a previously published fluoroscopy study were used to simulate realistic kinematics of daily activities using IMUs mounted to a six-degrees-of-freedom joint simulator. A model-based method involving extended Kalman filtering to derive rotational kinematics from inertial measurements was tested and compared against the ground truth simulator values. The algorithm demonstrated excellent accuracy (root-mean-square error ≤0.9°, maximum absolute error ≤3.2°) in estimating three-dimensional rotational knee kinematics during level walking. Although maximum absolute errors linked to stair descent and sit-to-stand-to-sit rose to 5.2° and 10.8°, respectively, root-mean-square errors peaked at 1.9° and 7.5°. This study hereby describes an accurate framework for evaluating the suitability of the underlying kinematic models and assumptions of an IMU-based motion analysis system, facilitating the future validation of analogous tools.
Soft tissue artefact affects the determination of skeletal kinematics. Thus, it is important to know the accuracy and limitations of kinematic parameters determined and modelled based on skin marker ...data. Here, the curvature angles, as well as the rotations of the lumbar and thoracic segments, of seven healthy subjects were determined in the sagittal plane using a skin marker set and compared to measurements taken in an open upright MRI scanner in order to understand the influence of soft tissue artefact at the back. The mean STA in the flexed compared to the extended positions were 10.2±6.1 mm (lumbar)/9.3±4.2 mm (thoracic) and 10.7±4.8 mm (lumbar)/9.2±4.9 mm (thoracic) respectively. A linear regression of the lumbar and thoracic curvatures between the marker-based measurements and MRI-based measurements resulted in coefficients of determination, R2, of 0.552 and 0.385 respectively. Skin marker measurements therefore allow for the assessment of changes in the lumbar and thoracic curvature angles, but the absolute values suffer from uncertainty. Nevertheless, this marker set appears to be suitable for quantifying lumbar and thoracic spinal changes between quasi-static whole body postural changes.
Video-fluoroscopic analysis can provide important insights for the evaluation of outcome and functionality after total knee arthroplasty, allowing the in vivo assessment of tibiofemoral kinematics ...without soft tissue artefacts. To enable measurement of the knee throughout activities of daily living such as gait, robotic systems like the moving fluoroscope have been developed that follow the knee movement and maintain the joint in front of the image intensifier. Since it is unclear whether walking while being accompanied by moving fluoroscope affects normal gait, the objective of this study was to investigate its influence on gait characteristics in healthy subjects. In addition, the impact of the motors' noise was analysed. By means of skin markers analysis (VICON MX system, Oxford Metrics Group, UK) and simultaneous measurement of ground reaction forces (Kistler force plates, Kistler, Switzerland), gait characteristics when walking with and without the moving fluoroscope as well as with and without ear protectors in combination with the moving fluoroscope, were obtained in young (n = 10, 24.5y ± 3.0y) and elderly (n = 9, 61.6y ± 5.3y) subjects during level gait and stair descent. Walking with the moving fluoroscope significantly decreased gait velocity in level gait and stair descent over the respective movement without the fluoroscope. Statistical analysis, including gait velocity as a covariate, resulted in no differences on the ground reaction force parameters. However, some kinematic parameters (ankle, knee and hip ranges of motion, minimal knee angle in late stance phase, maximal knee angles in stance and swing phase) seemed to be modified by the presence of the moving fluoroscope, but statistical comparison was limited due to velocity differences between the conditions. Wearing ear protectors to avoid the influence of motor sound during walking with the moving fluoroscope caused no significant difference. Walking with the moving fluoroscope has been shown to decrease gait velocity and small alterations in kinematic parameters were observed. Therefore, gait and movement alterations due to the moving fluoroscope cannot completely be excluded. However, based on the absence of differences in ground reaction force parameters (when adjusted for velocity within ANCOVA), as well as based on the comparable shape of the angular curves to the slow control condition, it can be concluded that changes in gait when walking with the moving fluoroscope are small, especially in comparison to natural slow walking. In order to allow assessment of joint replacement with the moving fluoroscope, including an understanding of the effects of joint pain, clinical analyses can only be compared to gait activities showing similarly reduced velocities. Importantly, the reduced gait speeds observed in this study are similar to those observed after total knee arthroplasty, suggesting that analyses in such subjects are appropriate. However, the moving fluoroscope would likely need to be optimized in order to detect natural gait characteristics at the higher gait velocities of healthy young subjects. The moving fluoroscope can be applied for comparisons between groups measured with the moving fluoroscope, but care should be taken when comparing data to subjects walking at self-selected speed without the moving fluoroscope.