Optically-pumped magnetometers (OPMs) have recently reached sensitivity levels required for magnetoencephalography (MEG). OPMs do not need cryogenics and can thus be placed within millimetres from ...the scalp into an array that adapts to the individual head size and shape, thereby reducing the distance from cortical sources to the sensors. Here, we quantified the improvement in recording MEG with hypothetical on-scalp OPM arrays compared to a 306-channel state-of-the-art SQUID array (102 magnetometers and 204 planar gradiometers).
We simulated OPM arrays that measured either normal (nOPM; 102 sensors), tangential (tOPM; 204 sensors), or all components (aOPM; 306 sensors) of the magnetic field. We built forward models based on magnetic resonance images of 10 adult heads; we employed a three-compartment boundary element model and distributed current dipoles evenly across the cortical mantle.
Compared to the SQUID magnetometers, nOPM and tOPM yielded 7.5 and 5.3 times higher signal power, while the correlations between the field patterns of source dipoles were reduced by factors of 2.8 and 3.6, respectively. Values of the field-pattern correlations were similar across nOPM, tOPM and SQUID gradiometers. Volume currents reduced the signals of primary currents on average by 10%, 72% and 15% in nOPM, tOPM and SQUID magnetometers, respectively. The information capacities of the OPM arrays were clearly higher than that of the SQUID array. The dipole-localization accuracies of the arrays were similar while the minimum-norm-based point-spread functions were on average 2.4 and 2.5 times more spread for the SQUID array compared to nOPM and tOPM arrays, respectively.
•We simulated on-scalp MEG arrays that measured normal or tangential field components.•On-scalp arrays showed higher signal powers and information content than SQUID array.•Point-spread functions of minimum-norm estimates were less spread in on-scalp arrays.•On-scalp MEG arrays offer clear benefits over SQUID arrays.
People are embedded in social interaction that shapes their brains throughout lifetime. Instead of emerging from lower-level cognitive functions, social interaction could be the default mode via ...which humans communicate with their environment. Should this hypothesis be true, it would have profound implications on how we think about brain functions and how we dissect and simulate them. We suggest that the research on the brain basis of social cognition and interaction should move from passive spectator science to studies including engaged participants and simultaneous recordings from the brains of the interacting persons.
Social interaction is central to human behavior and brain function. Hari et al. outline why brain imaging studies of social cognition should incorporate subjects’ active participation and discuss the promises and challenges involved in simultaneous brain imaging of interacting subjects.
To estimate the neural generators of magnetoencephalographic (MEG) signals, MEG data have to be co-registered with an anatomical image, typically an MR image. Optically-pumped magnetometers (OPMs) ...enable the construction of on-scalp MEG systems providing higher sensitivity and spatial resolution than conventional SQUID-based MEG systems. We present a co-registration method that can be applied to on-scalp MEG systems, regardless of the number of sensors. We apply a structured-light scanner to create a surface mesh of the subject's head and the sensor array, which we fit to the MR image. We quantified the reproducibility of the mesh and localised current dipoles with a phantom. Additionally, we measured somatosensory evoked fields (SEFs) to median nerve stimulation and compared the dipole positions between on-scalp and SQUID-based systems. The scanner reproduced the head surface with <1 mm error. Phantom dipoles were localised with 2.1 mm mean error. SEF dipoles corresponding to the P35m response for OPMs were well localised to the somatosensory cortex, while SQUID dipoles for two subjects were erroneously localised to the motor cortex. The developed co-registration method is inexpensive, fast and can easily be applied to on-scalp MEG. It is more convenient than traditional co-registration methods while also being more accurate.
The spatial resolution of magnetoencephalography (MEG) can be increased from that of conventional SQUID-based systems by employing on-scalp sensor arrays of e.g. optically-pumped magnetometers ...(OPMs). However, OPMs reach sufficient sensitivity for neuromagnetic measurements only when operated in a very low absolute magnetic field of few nanoteslas or less, usually not reached in a typical magnetically shielded room constructed for SQUID-based MEG. Moreover, field drifts affect the calibration of OPMs. Static and dynamic suppression of interfering fields is thus necessary for good-quality neuromagnetic measurements with OPMs. Here, we describe an on-scalp MEG system that utilizes OPMs and external compensation coils that provide static and dynamic shielding against ambient fields.
In a conventional two-layer magnetically shielded room, our coil system reduced the maximum remanent DC-field component within an 8-channel OPM array from 70 to less than 1 nT, enabling the sensors to operate in the sensitive spin exchange relaxation-free regime. When compensating field drifts below 4 Hz, a low-frequency shielding factor of 22 dB was achieved, which reduced the peak-to-peak drift from 1.3 to 0.4 nT and thereby the standard deviation of the sensor calibration from 1.7% to 0.5%. Without band-limiting the field that was compensated, a low-frequency shielding factor of 43 dB was achieved.
We validated the system by measuring brain responses to electric stimulation of the median nerve. With dynamic shielding and digital interference suppression methods, single-trial somatosensory evoked responses could be detected. Our results advance the deployment of OPM-based on-scalp MEG in lighter magnetic shields.
•We present an on-scalp MEG system based on optically-pumped magnetometers (OPMs).•The system uses active, dynamic shielding with external coils.•Active shielding reduces OPM calibration errors.•We demonstrate the operation of the system by measuring somatosensory responses.•Our solution facilitates deployment of OPM-based MEG in lighter magnetic shields.
Electrophysiological signals recorded intracranially show rich frequency content spanning from near‐DC to hundreds of hertz. Noninvasive electromagnetic signals measured with electroencephalography ...(EEG) or magnetoencephalography (MEG) typically contain less signal power in high frequencies than invasive recordings. Particularly, noninvasive detection of gamma‐band activity (>30 Hz) is challenging since coherently active source areas are small at such frequencies and the available imaging methods have limited spatial resolution. Compared to EEG and conventional SQUID‐based MEG, on‐scalp MEG should provide substantially improved spatial resolution, making it an attractive method for detecting gamma‐band activity. Using an on‐scalp array comprised of eight optically pumped magnetometers (OPMs) and a conventional whole‐head SQUID array, we measured responses to a dynamic visual stimulus known to elicit strong gamma‐band responses. OPMs had substantially higher signal power than SQUIDs, and had a slightly larger relative gamma‐power increase over the baseline. With only eight OPMs, we could obtain gamma‐activity source estimates comparable to those of SQUIDs at the group level. Our results show the feasibility of OPMs to measure gamma‐band activity. To further facilitate the noninvasive detection of gamma‐band activity, the on‐scalp OPM arrays should be optimized with respect to sensor noise, the number of sensors and intersensor spacing.
We discuss the importance of timing in brain function: how temporal dynamics of the world has left its traces in the brain during evolution and how we can monitor the dynamics of the human brain with ...non-invasive measurements. Accurate timing is important for the interplay of neurons, neuronal circuitries, brain areas and human individuals. In the human brain, multiple temporal integration windows are hierarchically organized, with temporal scales ranging from microseconds to tens and hundreds of milliseconds for perceptual, motor and cognitive functions, and up to minutes, hours and even months for hormonal and mood changes. Accurate timing is impaired in several brain diseases. From the current repertoire of non-invasive brain imaging methods, only magnetoencephalography (MEG) and scalp electroencephalography (EEG) provide millisecond time-resolution; our focus in this paper is on MEG. Since the introduction of high-density whole-scalp MEG/EEG coverage in the 1990s, the instrumentation has not changed drastically; yet, novel data analyses are advancing the field rapidly by shifting the focus from the mere pinpointing of activity hotspots to seeking stimulus- or task-specific information and to characterizing functional networks. During the next decades, we can expect increased spatial resolution and accuracy of the time-resolved brain imaging and better understanding of brain function, especially its temporal constraints, with the development of novel instrumentation and finer-grained, physiologically inspired generative models of local and network activity. Merging both spatial and temporal information with increasing accuracy and carrying out recordings in naturalistic conditions, including social interaction, will bring much new information about human brain function.
Magnetoencephalography and electroencephalography (M/EEG) measure the weak electromagnetic signals originating from neural currents in the brain. Using these signals to characterize and locate brain ...activity is a challenging task, as evidenced by several decades of methodological contributions. MNE, whose name stems from its capability to compute cortically-constrained minimum-norm current estimates from M/EEG data, is a software package that provides comprehensive analysis tools and workflows including preprocessing, source estimation, time–frequency analysis, statistical analysis, and several methods to estimate functional connectivity between distributed brain regions. The present paper gives detailed information about the MNE package and describes typical use cases while also warning about potential caveats in analysis. The MNE package is a collaborative effort of multiple institutes striving to implement and share best methods and to facilitate distribution of analysis pipelines to advance reproducibility of research. Full documentation is available at http://martinos.org/mne.
Display omitted
•The MNE software provides a complete pipeline for MEG and EEG data analysis.•MNE covers preprocessing, forward modeling, inverse methods, and visualization.•MNE supports advanced analysis: time-frequency, statistics, and connectivity.•MNE-Python enables fast and memory-efficient processing of large data sets.•MNE-Python is an open-source software supporting a collaborative development effort.
Sensory processing during development is important for the emerging cognitive skills underlying goal-directed behavior. Yet, it is not known how auditory processing in children is related to their ...cognitive functions. Here, we utilized combined magneto- and electroencephalographic (M/EEG) measurements in school-aged children (6-14y) to show that child auditory cortical activity at ∼250 ms after auditory stimulation predicts the performance in inhibition tasks. While unaffected by task demands, the amplitude of the left-hemisphere activation pattern was significantly correlated with the variability of behavioral response time. Since this activation pattern is typically not present in adults, our results suggest divergent brain mechanisms in adults and children for consistent performance in auditory-based cognitive tasks. This difference can be explained as a shift in cortical resources for cognitive control from sensorimotor associations in the auditory cortex of children to top–down regulated control processes involving (pre)frontal and cingulate areas in adults.
Brain-computer interfaces (BCI) can be designed with several feedback modalities. To promote appropriate brain plasticity in therapeutic applications, the feedback should guide the user to elicit the ...desired brain activity and preferably be similar to the imagined action. In this study, we employed magnetoencephalography (MEG) to measure neurophysiological changes in healthy subjects performing motor imagery (MI) -based BCI training with two different feedback modalities. The MI-BCI task used in this study lasted 40-60 min and involved imagery of right- or left-hand movements. 8 subjects performed the task with visual and 14 subjects with proprioceptive feedback. We analysed power changes across the session at multiple frequencies in the range of 4-40 Hz with a generalized linear model to find those frequencies at which the power increased significantly during training. In addition, the power increase was analysed for each gradiometer, separately for alpha (8-13 Hz), beta (14-30 Hz) and gamma (30-40 Hz) bands, to find channels showing significant linear power increase over the session. These analyses were applied during three different conditions: rest, preparation, and MI. Visual feedback enhanced the amplitude of mainly high beta and gamma bands (24-40 Hz) in all conditions in occipital and left temporal channels. During proprioceptive feedback, in contrast, power increased mainly in alpha and beta bands. The alpha-band enhancement was found in multiple parietal, occipital, and temporal channels in all conditions, whereas the beta-band increase occurred during rest and preparation mainly in the parieto-occipital region and during MI in the parietal channels above hand motor regions. Our results show that BCI training with proprioceptive feedback increases the power of sensorimotor rhythms in the motor cortex, whereas visual feedback causes mainly a gamma-band increase in the visual cortex. MI-BCIs should involve proprioceptive feedback to facilitate plasticity in the motor cortex.
In MEG and EEG studies, the accuracy of the head digitization impacts the co-registration between functional and structural data. The co-registration is one of the major factors that affect the ...spatial accuracy in MEG/EEG source imaging. Precisely digitized head-surface (scalp) points do not only improve the co-registration but can also deform a template MRI. Such an individualized-template MRI can be used for conductivity modeling in MEG/EEG source imaging if the individual's structural MRI is unavailable. Electromagnetic tracking (EMT) systems (particularly Fastrak, Polhemus Inc., Colchester, VT, USA) have been the most common solution for digitization in MEG and EEG. However, they may occasionally suffer from ambient electromagnetic interference which makes it challenging to achieve (sub-)millimeter digitization accuracy. The current study-(i) evaluated the performance of the Fastrak EMT system under different conditions in MEG/EEG digitization, and (ii) explores the usability of two alternative EMT systems (Aurora, NDI, Waterloo, ON, Canada; Fastrak with a short-range transmitter) for digitization. Tracking fluctuation, digitization accuracy, and robustness of the systems were evaluated in several test cases using test frames and human head models. The performance of the two alternative systems was compared against the Fastrak system. The results showed that the Fastrak system is accurate and robust for MEG/EEG digitization if the recommended operating conditions are met. The Fastrak with the short-range transmitter shows comparatively higher digitization error if digitization is not carried out very close to the transmitter. The study also evinces that the Aurora system can be used for MEG/EEG digitization within a constrained range; however, some modifications would be required to make the system a practical and easy-to-use digitizer. Its real-time error estimation feature can potentially improve digitization accuracy.