    Tayari, Nassim; Obels, Jiri; Kobus, Thiele; Scheenen, Tom W.J.; Heerschap, Arend

    Magnetic resonance in medicine, 20/May , Letnik: 81, Številka: 5
    Journal Article

    Purpose Quality control (QC) is a prerequisite for clinical MR spectroscopic imaging (MRSI) to avoid that bad spectra hamper data interpretation. The aim of this work was to present a simple automatic QC for prostate 1H MRSI that can handle data obtained with different commonly used pulse sequences, echo times, field strengths, and MR platforms. Methods A QC method was developed with a ratio (Qratio) where the numerator and the denominator are functions of several signal heights, logically combined for their positive or negative contribution to spectral quality. This Qratio was tested on 4 data sets obtained at 1.5, 3, and 7T, with and without endorectal coil and different localization sequences and echo times. Spectra of 25,248 voxels in 26 prostates were labeled as acceptable or unacceptable by MRS experts as gold standard. A threshold value was determined for Qratio from a subset of voxels, labeled in consensus by 4 experts, for an optimal accuracy to separate spectra. Results Applying this Qratio threshold to the remaining test voxels, an automatic separation of good and bad spectra was possible with an accuracy of 0.88, similar to manual separation between the 2 classes. Qratio values were used to generate maps representing spectral quality on a binary or continuous scale. Conclusion Automated QC of prostate 1H MRSI by Qratio is fast, simple, easily transferable and more practical than supervised feature extraction methods and therefore easy to integrate into different clinical MR systems. Moreover, quality maps can be generated to read the reliability of spectra in each voxel.