  • Validity and reliability of...
    Balsalobre-Fernández, Carlos; Marchante, David; Muñoz-López, Mario; Jiménez, Sergio L.

    Journal of sports sciences, 01/2018, Letnik: 36, Številka: 1
    Journal Article

    The purpose of this study was to analyse the validity and reliability of a novel iPhone app (named: PowerLift) for the measurement of mean velocity on the bench-press exercise. Additionally, the accuracy of the estimation of the 1-Repetition maximum (1RM) using the load-velocity relationship was tested. To do this, 10 powerlifters (Mean (SD): age = 26.5 ± 6.5 years; bench press 1RM · kg −1  = 1.34 ± 0.25) completed an incremental test on the bench-press exercise with 5 different loads (75-100% 1RM), while the mean velocity of the barbell was registered using a linear transducer (LT) and Powerlift. Results showed a very high correlation between the LT and the app (r = 0.94, SEE = 0.028 m · s −1 ) for the measurement of mean velocity. Bland-Altman plots (R 2  = 0.011) and intraclass correlation coefficient (ICC = 0.965) revealed a very high agreement between both devices. A systematic bias by which the app registered slightly higher values than the LT (P < 0.05; mean difference (SD) between instruments = 0.008 ± 0.03 m · s −1 ). Finally, actual and estimated 1RM using the app were highly correlated (r = 0.98, mean difference (SD) = 5.5 ± 9.6 kg, P < 0.05). The app was found to be highly valid and reliable in comparison with a LT. These findings could have valuable practical applications for strength and conditioning coaches who wish to measure barbell velocity in the bench-press exercise.