Pejović, Milić; Živanović, Emilija; Živanović, Miloš
Radiation protection dosimetry, 08/2020, Letnik: 190, Številka: 1Journal Article
Abstract This paper presents experimental results of dynamic breakdown voltage and delay response as functions of gamma ray air kerma rate for xenon-filled tube at 2.7 mbar pressure. Gamma ray air kerma rate range was considered from 123 nGy h–1 up to 12.3 mGy h–1 in order to investigate the possibility of the application of this tube in gamma radiation dosimetry. It was shown that the variations of the above-mentioned parameters are considerable up to the dose rate of 1.23 μGy h–1, which points to the possibility for application in small dose rate gamma ray dosimetry. Physical processes that make dominant impact to dynamic breakdown voltage and delay response during xenon-filled tube irradiation are also discussed in the paper.