  • Impact of tip design and th...
    Yamaguchi, Junji; Takigawa, Masateru; Goya, Masahiko; Martin, Claire; Amemiya, Miki; Yamamoto, Tasuku; Nishimura, Takuro; Nakamura, Rena; Shirai, Yasuhiro; Tao, Susumu; Miyazaki, Shinsuke; Takahashi, Yoshihide; Sasano, Tetsuo

    Journal of interventional cardiac electrophysiology, 06/2023, Letnik: 66, Številka: 4
    Journal Article

    Background The FlexAbility™ SE catheter has a laser-cut 8Fr 4-mm flexible tip irrigated through laser-cut kerfs with a thermocouple 0.3 mm from the distal end. The TactiCath™ SE catheter has an 8Fr 3.5-mm tip and 6-irrigation port with a thermocouple 2.67 mm proximal to the tip. We investigated the impact of these differences on the efficacy and safety of radiofrequency (RF) applications. Methods RF applications at a range of powers (20 W, 30 W, and 40 W), contact forces (5 g, 15 g, and 25 g), and durations (10–60 s) using perpendicular/parallel catheter orientation were performed in excised porcine hearts. Lesion characteristics and incidence of steam pops were compared. Results A total of 540 lesions were examined. The FlexAbility™ SE catheter produced smaller lesion depths (4.0 mm vs. 4.4 mm, p  = 0.014 at 20 W; 4.6 mm vs. 5.6 mm, p  = 0.015 at 30 W), surface areas (22.7mm 2 vs. 29.2mm 2 at 20 W, p  = 0.005; 23.2mm 2 vs. 28.7mm 2 , p  = 0.009 at 30 W), and volumes (126.1mm 3 vs. 175.1mm 3 , p  = 0.018 at 20 W; 183.2mm 3 vs. 304.3mm 3 , p  = 0.002 at 30 W) with perpendicular catheter placement. However, no differences were observed with parallel catheter placement. Steam-pops were significantly less frequently observed with the FlexAbility™ SE catheter (4% vs. 22%, p  < 0.001) irrespective of catheter direction to the tissue. Multivariate analysis showed that use of the TactiCath™ SE catheter, power ≥ 40 W, contact force ≥ 25 g, RF duration > 30 s, parallel angle, and impedance drop ≥ 20Ω were significantly associated with occurrence of steam-pops. Conclusions The FlexAbility™ SE catheter reduced the risk of steam-pops but produced smaller lesions with perpendicular catheter placement compared to the TactiCath™ SE catheter.