DIKUL - logo
E-viri
Celotno besedilo
Recenzirano
  • Stability of pacing indices...
    Farid, Talha A.; Omer, Mohamed A.; Gosch, Kensey; Moser, Ashley; Austin, Bethany; Magalski, Anthony; Wimmer, Alan P.

    Journal of interventional cardiac electrophysiology, 06/2017, Letnik: 49, Številka: 1
    Journal Article

    Background A significant minority of cardiac transplant patients require permanent pacemaker (PPM) implant, primarily for sinus node dysfunction. The stability of pacing indices has not been determined in this unique patient population, and data regarding ongoing need for pacing are limited. Methods Pacing indices (sensing, threshold, and impedance) as well as the percentage of time patients required pacing were recorded, from 30 cardiac transplant patients that underwent PPM implant, over 1 year of follow-up. Repeated measure ANOVA (analysis of variance) was used to compare pacing indices and the percentage of time patients required pacing in each cardiac chamber (right atrium (RA) and right ventricle (RV)) and at different time points. Results There was no difference in sensing among the follow-up time points ( p  = 0.9). Thresholds at 3 months were significantly higher compared to the day of implant ( p  = 0.005) and the day after implant ( p  = 0.03). Impedances at implant were significantly higher compared to day 1 ( p  < 0.001), 3 months ( p  < 0.003), and 12 months ( p  < 0.001) post-implant. The mean percentage of RA pacing was 85 ± 6% the day after implant, 74 ± 6% at 3 months, and 80 ± 6% at 1 year ( p  = 0.17). Conclusion In cardiac transplant patients, pacing impedances decrease and thresholds trend up in short-term follow-up, but subsequent sensing, threshold, and impedance remain stable at 1 year. This is comparable to the pattern observed among noncardiac transplant PPM recipients. The atrial pacing percentage was stable over 1 year, suggesting continued relative sinus node dysfunction.