Purpose
Spatial distortions in magnetic resonance imaging (MRI) are mainly caused by inhomogeneities of the static magnetic field, nonlinearities in the applied gradients, and tissue‐specific ...magnetic susceptibility variations. These factors may significantly alter the geometrical accuracy of the reconstructed MR image, thus questioning the reliability of MRI for guidance in image‐guided radiation therapy. In this work, we quantified MRI spatial distortions and created a quantitative model where different sources of distortions can be separated. The generated model was then integrated into a four‐dimensional (4D) computational phantom for simulation studies in MRI‐guided radiation therapy at extra‐cranial sites.
Methods
A geometrical spatial distortion phantom was designed in four modules embedding laser‐cut PMMA grids, providing 3520 landmarks in a field of view of (345 × 260 × 480) mm3. The construction accuracy of the phantom was verified experimentally. Two fast MRI sequences for extra‐cranial imaging at 1.5 T were investigated, considering axial slices acquired with online distortion correction, in order to mimic practical use in MRI‐guided radiotherapy. Distortions were separated into their sources by acquisition of images with gradient polarity reversal and dedicated susceptibility calculations. Such a separation yielded a quantitative spatial distortion model to be used for MR imaging simulations. Finally, the obtained spatial distortion model was embedded into an anthropomorphic 4D computational phantom, providing registered virtual CT/MR images where spatial distortions in MRI acquisition can be simulated.
Results
The manufacturing accuracy of the geometrical distortion phantom was quantified to be within 0.2 mm in the grid planes and 0.5 mm in depth, including thickness variations and bending effects of individual grids. Residual spatial distortions after MRI distortion correction were strongly influenced by the applied correction mode, with larger effects in the trans‐axial direction. In the axial plane, gradient nonlinearities caused the main distortions, with values up to 3 mm in a 1.5 T magnet, whereas static field and susceptibility effects were below 1 mm. The integration in the 4D anthropomorphic computational phantom highlighted that deformations can be severe in the region of the thoracic diaphragm, especially when using axial imaging with 2D distortion correction. Adaptation of the phantom based on patient‐specific measurements was also verified, aiming at increased realism in the simulation.
Conclusions
The implemented framework provides an integrated approach for MRI spatial distortion modeling, where different sources of distortion can be quantified in time‐dependent geometries. The computational phantom represents a valuable platform to study motion management strategies in extra‐cranial MRI‐guided radiotherapy, where the effects of spatial distortions can be modeled on synthetic images in a virtual environment.
X-ray imaging with gaseous detectors using the VMM3a and the SRS Scharenberg, L.; Bortfeldt, J.; Brunbauer, F. ...
Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment,
09/2021, Volume:
1011
Journal Article
Peer reviewed
Open access
The integration of the VMM3a Application-Specific Integrated Circuit (ASIC) into RD51’s Scalable Readout System (SRS) provides a versatile tool for the readout of Micro-Pattern Gaseous Detectors ...(MPGDs). With its self-triggered high-rate readout, its analogue part that allows to get information on the deposited energy in the detector, and its so-called neighbouring-logic that allows to recover information on the charge distribution, this new system has features of particular interest for digital X-ray imaging. In the present article, we want to emphasise the capabilities of VMM3a/SRS by presenting results of X-ray imaging studies. We will highlight the advantages on the energy and the spatial resolution provided by the neighbouring-logic. In the first part, we focus on spatial resolution studies. We show how segmented readout structures introduce a repeating pattern in the distribution of the reconstructed positions (using the centre-of-gravity method) and how this behaviour can be mitigated with the neighbouring-logic. As part of these studies, we explore as well an alternative position reconstruction algorithm. In the second part of the article, we present the energy resolution studies.
Timing performance of a multi-pad PICOSEC-Micromegas detector prototype Aune, S.; Bortfeldt, J.; Brunbauer, F. ...
Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment,
03/2021, Volume:
993, Issue:
C
Journal Article
Peer reviewed
Open access
The multi-pad PICOSEC-Micromegas is an improved detector prototype with a segmented anode, consisting of 19 hexagonal pads. Detailed studies are performed with data collected in a muon beam over four ...representative pads. We demonstrate that such a device, scalable to a larger area, provides excellent time resolution and detection efficiency. As expected from earlier single-cell device studies, we measure a time resolution of approximately 25 picoseconds for charged particles hitting near the anode pad centres, and up to 30 picoseconds at the pad edges. Here, we study in detail the effect of drift gap thickness non-uniformity on the timing performance and evaluate impact position based corrections to obtain a uniform timing response over the full detector coverage.
Magnetic-resonance linear-accelerator (MR-LINAC) systems integrating in-room magnetic-resonance-imaging (MRI) guidance are a currently emerging technology. Such systems address the need to provide ...frequent imaging at optimal soft-tissue contrast for treatment guidance. However, the use of MRI-guidance in radiotherapy should address imaging-related spatial distortions, which may hinder accurate geometrical characterization of the treatment site. Since spatial encoding relies on well-defined magnetic fields, accurate modeling of the magnetic field alterations due to Formula: see text-inhomogeneities, gradient nonlinearities, and susceptibilities is needed. In this work, the modeling of susceptibility induced distortions is considered. Dedicated susceptibility measurements are reported, aiming at extending the characterization of different tissues for MRI-guided extra-cranial radiotherapy applications. A digital 4D anthropomorphic phantom, providing time-resolved anatomical changes due to breathing, is exploited as reference anatomy to quantify spatial distortions due to variations in tissue susceptibility. Sub-millimeter values can be attributed to susceptibility-induced distortions, with maximum values up to 2.3 mm at a gradient strength of 5 mT m
. Improvements in susceptibility simulation for extra-cranial sites are shown when including specifically the contributions from lung, liver and muscular tissues.
The PICOSEC Micromegas detector can time the arrival of Minimum Ionizing Particles with a sub-25 ps precision. A very good timing resolution in detecting single photons is also demonstrated in laser ...beams. The PICOSEC timing resolution is determined mainly by the drift field. The arrival time of the signal and the timing resolution vary with the size of the pulse amplitude.
Detailed simulations based on GARFIELD++ reproduce the experimental PICOSEC timing characteristics. This agreement is exploited to identify the microscopic physical variables, which determine the observed timing properties. In these studies, several counter-intuitive observations are made for the behavior of such microscopic variables. In order to gain insight on the main physical mechanisms causing the observed behavior, a phenomenological model is constructed and presented. The model is based on a simple mechanism of “time-gain per interaction” and it employs a statistical description of the avalanche evolution. It describes quantitatively the dynamical and statistical properties of the microscopic quantities, which determine the PICOSEC timing characteristics, in excellent agreement with the simulations. In parallel, it offers phenomenological explanations for the behavior of these microscopic variables. The formulae expressing this model can be used as a tool for fast and reliable predictions, provided that the input parameter values (e.g. drift velocities) are known for the considered operating conditions.
This contribution describes the PICOSEC-Micromegas detector which achieves a time resolution below 25ps. In this device the passage of a charged particle produces Cherenkov photons in a radiator, ...which then generate electrons in a photocathode and these photoelectrons enter a two-stage Micromegas with a reduced drift region and a typical anode region. The results from single-channel prototypes (demonstrating a time resolution of 24ps for minimum ionizing particles, and 76ps for single photoelectrons), the understanding of the detector in terms of detailed simulations and a phenomenological model, the issues of robustness and how they are tackled, and preliminary results from a multi-channel prototype are presented (demonstrating that a timing resolution similar to that of the single-channel device is feasible for all points across the area covered by a multi-channel device).
Floating strip Micromegas are high-accuracy and discharge insensitive gaseous detectors, able to track single particles at fluxes of 7MHz/cm2 with 100μm resolution. We developed low-material-budget ...detectors with one-dimensional strip readout, suitable for tracking at highest particle rates as encountered in medical ion transmission radiography or inner tracker applications. Recently we additionally developed Kapton-based floating strip Micromegas with two-dimensional strip readout, featuring an overall thickness of 0.011 X0.
These detectors were tested in high-rate proton and carbon-ion beams at the tandem accelerator in Garching and the Heidelberg Ion-Beam Therapy Center, operated with an optimized Ne:CF4 gas mixture. By coupling the Micromegas detectors to a new scintillator based range detector, ion transmission radiographies of PMMA and tissue-equivalent phantoms were acquired. The range detector with 18 layers is read out via wavelength shifting fibers, coupled to a multi-anode photomultiplier.
We present the performance of the Micromegas detectors with respect to timing and single plane track reconstruction using the μTPC method. We discuss the range resolution of the scintillator range telescope and present the image reconstruction capabilities of the combined system.