. The Monte Carlo simulation software is a valuable tool in radiation therapy, in particular to achieve the needed accuracy in the dose evaluation for the treatment plans optimisation. The current ...challenge in this field is the time reduction to open the way to many clinical applications for which the computational time is an issue. In this manuscript we present an innovative GPU-accelerated Monte Carlo software for dose valuation in electron and photon based radiotherapy, developed as an update of the FRED (Fast paRticle thErapy Dose evaluator) software.
. The code transports particles through a 3D voxel grid, while scoring their energy deposition along their trajectory. The models of electromagnetic interactions in the energy region between 1 MeV-1 GeV available in literature have been implemented to efficiently run on GPUs, allowing to combine a fast tracking while keeping high accuracy in dose assessment. The FRED software has been bench-marked against state-of-art full MC (FLUKA, GEANT4) in the realm of two different radiotherapy applications: Intra-Operative Radio Therapy and Very High Electron Energy radiotherapy applications.
. The single pencil beam dose-depth profiles in water as well as the dose map computed on non-homogeneous phantom agree with full-MCs at 2% level, observing a gain in processing time from 200 to 5000.
. Such performance allows for computing a plan with electron beams in few minutes with an accuracy of ∼%, demonstrating the FRED potential to be adopted for fast plan re-calculation in photon or electron radiotherapy applications.
The high dose conformity and healthy tissue sparing achievable in Particle Therapy when using C ions calls for safety factors in treatment planning, to prevent the tumor under-dosage related to the ...possible occurrence of inter-fractional morphological changes during a treatment. This limitation could be overcome by a range monitor, still missing in clinical routine, capable of providing on-line feedback. The Dose Profiler (DP) is a detector developed within the INnovative Solution for In-beam Dosimetry in hadronthErapy (INSIDE) collaboration for the monitoring of carbon ion treatments at the CNAO facility (Centro Nazionale di Adroterapia Oncologica) exploiting the detection of charged secondary fragments that escape from the patient. The DP capability to detect inter-fractional changes is demonstrated by comparing the obtained fragment emission maps in different fractions of the treatments enrolled in the first ever clinical trial of such a monitoring system, performed at CNAO. The case of a CNAO patient that underwent a significant morphological change is presented in detail, focusing on the implications that can be drawn for the achievable inter-fractional monitoring DP sensitivity in real clinical conditions. The results have been cross-checked against a simulation study.
Tumour control is performed in particle therapy using particles and ions, whose high irradiation precision enhances the effectiveness of the treatment, while sparing the healthy tissue surrounding ...the target volume. Dose range monitoring devices using photons and charged particles produced by the beam interacting with the patient's body have already been proposed, but no attempt has been made yet to exploit the detection of the abundant neutron component. Since neutrons can release a significant dose far away from the tumour region, precise measurements of their flux, production energy and angle distributions are eagerly sought in order to improve the treatment planning system (TPS) software. It will thus be possible to predict not only the normal tissue toxicity in the target region, but also the risk of late complications in the whole body. The aforementioned issues underline the importance of an experimental effort devoted to the precise characterisation of neutron production, aimed at the measurement of their abundance, emission point and production energy. The technical challenges posed by a neutron detector aimed at high detection efficiency and good backtracking precision are addressed within the MONDO (monitor for neutron dose in hadrontherapy) project, whose main goal is to develop a tracking detector that can target fast and ultrafast neutrons. A full reconstruction of two consecutive elastic scattering interactions undergone by the neutrons inside the detector material will be used to measure their energy and direction. The preliminary results of an MC simulation performed using the FLUKA software are presented here, together with the DSiPM (digital SiPM) readout implementation. New detector readout implementations specifically tailored to the MONDO tracker are also discussed, and the neutron detection efficiency attainable with the proposed neutron tracking strategy are reported.
The high dose conformity and healthy tissue sparing achievable in Particle Therapy when using C ions calls for safety factors in treatment planning, to prevent the tumor under-dosage related to the ...possible occurrence of inter-fractional morphological changes during a treatment. This limitation could be overcome by a range monitor, still missing in clinical routine, capable of providing on-line feedback. The Dose Profiler (DP) is a detector developed within the INnovative Solution for In-beam Dosimetry in hadronthErapy (INSIDE) collaboration for the monitoring of carbon ion treatments at the CNAO facility (Centro Nazionale di Adroterapia Oncologica) exploiting the detection of charged secondary fragments that escape from the patient. The DP capability to detect inter-fractional changes is demonstrated by comparing the obtained fragment emission maps in different fractions of the treatments enrolled in the first ever clinical trial of such a monitoring system, performed at CNAO. The case of a CNAO patient that underwent a significant morphological change is presented in detail, focusing on the implications that can be drawn for the achievable inter-fractional monitoring DP sensitivity in real clinical conditions. The results have been cross-checked against a simulation study.
The MOnitor for Neutron Dose in hadrOntherapy (MONDO) project addresses the technical challenges posed by a neutron tracker detector aiming for a high detection efficiency and a good backtracking ...precision. The project aims to develop a tracking device capable of fully reconstructing the four momentum of fast and ultrafast secondary neutrons produced, e.g., in particle therapy (PT) treatments or in other physical processes. The MONDO tracker uses, as active material, squared scintillating fibers readout by dedicated CMOS-based digital single-photon avalanche diode (SPAD) array sensors. The expected light output, when operating in neutron monitoring applications, was experimentally evaluated in order to optimize the design of the MONDO detector readout. A small detector prototype (4 × 4 × 4.8 cm) has been built and tested at a test beam facility. The detection capabilities have been measured using a traditional photomultiplier (PMT) and a particle beam of 450-MeV electrons crossing a single layer of fibers. The observed number of photoelectrons in this case is (7.2 ± 1.4). A detector prototype was also tested with an SPAD-based SBAM (SPAD-Based Acquisition readout for MONDO experiment) sensor (SPADnet-I) to study the tracking performances. SBAM is a novel sensor developed to match the need of high single-photon detection efficiency and high spatial resolution and compactness. The sensor expected performance is discussed in view of an operation tailored for PT applications. In this contribution, we also report the results of a simulation performed to optimize the full MONDO detector layout.
Radio-Guided Surgery (RGS) is a nuclear medicine technique to support the surgeon during surgery towards a complete tumor resection. It is based on intraoperative detection of radiation emitted by a ...radio-pharmaceutical that bounds selectively to tumoral cells. In the past years, an approach that exploits β− emitting radiotracers has been pursued to overtake some limitations of the traditional RGS based on γ emission.
A particle detector dedicated to this application, demonstrating very high efficiency to β− particles and remarkable transparency to photons, has been thus developed. As a by-product, its characteristics suggested the possibility to utilize it with β+ emitting sources, more commonly in use in nuclear medicine. In this paper, performances of such detector on 18F liquid sources are estimated by means of Monte Carlo simulations (MC) and laboratory measurements. The experimental setup with a 18F saline solution comprised a “positron signal” spot (a 7 × 10 mm cylinder representing the tumor residual), and a surrounding “far background” volume, that represented for the detector an almost isotropic source of annihilation photons.
Experimental results show good agreement with MC predictions, thus confirming the expected performances of the detector with 18F, and the validity of the developed MC simulation as a tool to predict the gamma background determined by a diffuse source of annihilation photons.
•Radio Guided Surgery (RGS) based on β+ radiation extends its range of application.•This novel β-probe can detect 18F betas, with low annihilation photon background.•A simulation has been developed and validated to foresee the impact of background.
Radioguided surgery (RGS) is a medical practice which thanks to a radiopharmaceutical tracer and a probe allows the surgeon to identify tumor residuals up to a millimetric resolution in real-time. ...The employment of β− emitters, instead of γ or β+, reduces background from healthy tissues, administered activity to the patient, and medical exposure. In a previous work the possibility of using a CMOS Imager (Aptina MT9V011), initially designed for visible light imaging, to detect β− from 90Y or 90Sr sources has been established. Because of its possible application as counting probe in RGS, the performances of MT9V011 in clinical-like conditions were studied.11This work financed by Istituto Nazionale Fisica Nucleare, Italy, project CHIR2, and partially supported by Universitá degli Studi di Perugia, Italy , Fondo Ricerca di Base 2017, project SEISIPO.
Through horizontal scans on a collimated 90Sr source of different sizes (1, 3, 5, 7 mm), we have determined relationships between scan fit parameters and the source dimension, namely A quadratic correlation and a linear dependency of, respectively, signal integrated over scan interval, and maximum signal against source diameter, are determined. Horizontal scan measurements on a source, interposing collimators of different size, aim to determine relationships or correlations between scan fit parameters and source dimension. A quadratic correlation and a linear dependency of, respectively, signal integrated over scan interval, and maximum signal against source diameter are determined.
In order to get closer to clinical conditions, agar–agar phantoms containing 90Y with different dimensions and activities were prepared. A 90Y phantom is characterized by a central spot and a ring all around, for simulating both signal (tumor) and background (surrounding healthy tissue). The relationship found between scan maximum and 90Sr source diameter is then exploited to extract the concentration ratio between spot and external ring of the 90Y phantom. This observable, defined as the ratio between the tumor and the nearby healthy tissues uptake simulates the Tumor-to-Non-tumor Ratio (TNR). With the aim of evaluating the sensor’s ability to discriminate signal from background relying on the significance parameter, a further 90Y phantom, featuring a well-known and clinical-like activity will mimic the signal only condition. This result is used to extrapolate to different source sizes, after having estimated the background for various TNR. The obtained significance values suggest that the MT9V011 sensor is capable of distinguishing a signal from an estimated background, depending on the interplay among TNR, acquisition time and tumor diameter.
•A CMOS imager has been used as sensor for beta- emission from isotopes (90Y) of interest in Radioguided surgery.•Using a position scan a correlation between the source dimension and the shape of the response has been found.•The sensitivity of the sensor to tumor detection has been determined by studying the interplay among TNR, acquisition time and tumor diameter.
•The Dose Profiler is a charged fragment tracker designed for range monitoring in particle therapy.•The detector design, carefully optimized to operate in clinical environment, is described.•The ...characterization measurements have been performed using different experimental setup.•The obtained performances are suitable for range monitoring application.
Particle therapy (PT) can exploit heavy ions (such as He, C or O) to enhance the treatment efficacy, profiting from the increased Relative Biological Effectiveness and Oxygen Enhancement Ratio of these projectiles with respect to proton beams. To maximise the gain in tumor control probability a precise online monitoring of the dose release is needed, avoiding unnecessary large safety margins surroundings the tumor volume accounting for possible patient mispositioning or morphological changes with respect to the initial CT scan. The Dose Profiler (DP) detector, presented in this manuscript, is a scintillating fibres tracker of charged secondary particles (mainly protons) that will be operating during the treatment, allowing for an online range monitoring. Such monitoring technique is particularly promising in the context of heavy ions PT, in which the precision achievable by other techniques based on secondary photons detection is limited by the environmental background during the beam delivery. Developed and built at the SBAI department of “La Sapienza”, within the INSIDE collaboration and as part of a Centro Fermi flagship project, the DP is a tracker detector specifically designed and planned for clinical applications inside a PT treatment room. The DP operation in clinical like conditions has been tested with the proton and carbon ions beams of Trento proton-therapy center and of the CNAO facility. In this contribution the detector performances are presented, in the context of the carbon ions monitoring clinical trial that is about to start at the CNAO centre.