  • Ratheesh, Ajay; Rankin, John; Costa, Enrico; Ettore Del Monte; Alessandro Di Marco; Fabiani, Sergio; Fabio La Monaca; Muleri, Fabio; Rubini, Alda; Soffitta, Paolo; Baldini, Luca; Minuti, Massimo; Pinchera, Michele; Sgrò, Carmelo

    arXiv.org, 07/2023
    Paper, Journal Article

    In this work, we measured the polarization properties of the X-rays emitted from the X-ray tubes, which were used during the calibration of the instrument onboard Imaging X-ray Polarimetry Explorer (IXPE). X-ray tubes are used as a source of unpolarized X-rays to calibrate the response of the gas pixel detectors to unpolarized radiation. However, even though the characteristic fluorescent emission lines are unpolarized, continuum bremsstrahlung emission can be polarized based on the geometry of the accelerated electrons and emitted photons. Hence, characterizing the contribution of polarized X-rays from bremsstrahlung emission is of interest, also for future measurements. We find that when accelerated electrons are parallel to the emitted photons, the bremsstrahlung emission is unpolarized, and when they are perpendicular, the polarization increases with energy, as expected from the theoretical predictions. A comparison with the theoretical predictions is also shown.