  • Automated quality control f...
    Bastiani, Matteo; Cottaar, Michiel; Fitzgibbon, Sean P.; Suri, Sana; Alfaro-Almagro, Fidel; Sotiropoulos, Stamatios N.; Jbabdi, Saad; Andersson, Jesper L.R.

    NeuroImage (Orlando, Fla.), 01/2019, Volume: 184
    Journal Article

    Diffusion MRI data can be affected by hardware and subject-related artefacts that can adversely affect downstream analyses. Therefore, automated quality control (QC) is of great importance, especially in large population studies where visual QC is not practical. In this work, we introduce an automated diffusion MRI QC framework for single subject and group studies. The QC is based on a comprehensive, non-parametric approach for movement and distortion correction: FSL EDDY, which allows us to extract a rich set of QC metrics that are both sensitive and specific to different types of artefacts. Two different tools are presented: QUAD (QUality Assessment for DMRI), for single subject QC and SQUAD (Study-wise QUality Assessment for DMRI), which is designed to enable group QC and facilitate cross-studies harmonisation efforts. •Two tools to automatically perform QC of diffusion MRI data.•Automated generation of single subject reports for visual inspection and database.•Group databases and reports allow to compare subjects within and between studies.•Categorical and continuous variables can be used to update the reports.