The development of dedicated positron emission tomography scanners is an active area of research, especially aiming at the improvement of lesion detection and in support of cancer treatment and ...management. Recently, dedicated Positron Emission Tomography (PET) systems with different configurations for specific organs have been developed for improving detection effectiveness. Open geometries are always subject to distortion and artifacts in the reconstructed images. Therefore, the aim of this work is to determine the optimal geometry for a novel cardiac PET system that will be developed by our team, and determine the time resolution needed to achieve reasonable image quality for the chosen geometry. The proposed geometries consist of 36 modules. These modules are arranged in two sets of two plates, each one with different configurations. We performed Monte Carlo simulations with different TOF resolutions, in order to test the image quality improvement in each case. Our results show, as expected, that increasing TOF resolution reduces distortion and artifact effects. We can conclude that a TOF resolution of the order of 200 ps is needed to reduce the artifacts, to acceptable levels, generated in the simulated cardiac-PET open geometries.
Purpose:
The authors have developed a trimodal PET/SPECT/CT scanner for small animal imaging. The gamma ray subsystems are based on monolithic crystals coupled to multianode photomultiplier tubes ...(MA-PMTs), while computed tomography (CT) comprises a commercially available microfocus x-ray tube and a CsI scintillator 2D pixelated flat panel x-ray detector. In this study the authors will report on the design and performance evaluation of the multimodal system.
Methods:
X-ray transmission measurements are performed based on cone-beam geometry. Individual projections were acquired by rotating the x-ray tube and the 2D flat panel detector, thus making possible a transaxial field of view (FOV) of roughly 80 mm in diameter and an axial FOV of 65 mm for the CT system. The single photon emission computed tomography (SPECT) component has a dual head detector geometry mounted on a rotating gantry. The distance between the SPECT module detectors can be varied in order to optimize specific user requirements, including variable FOV. The positron emission tomography (PET) system is made up of eight compact modules forming an octagon with an axial FOV of 40 mm and a transaxial FOV of 80 mm in diameter. The main CT image quality parameters (spatial resolution and uniformity) have been determined. In the case of the SPECT, the tomographic spatial resolution and system sensitivity have been evaluated with a99mTc solution using single-pinhole and multi-pinhole collimators. PET and SPECT images were reconstructed using three-dimensional (3D) maximum likelihood and ordered subset expectation maximization (MLEM and OSEM) algorithms developed by the authors, whereas the CT images were obtained using a 3D based FBP algorithm.
Results:
CT spatial resolution was 85μm while a uniformity of 2.7% was obtained for a water filled phantom at 45 kV. The SPECT spatial resolution was better than 0.8 mm measured with a Derenzo-like phantom for a FOV of 20 mm using a 1-mm pinhole aperture collimator. The full width at half-maximum PET radial spatial resolution at the center of the field of view was 1.55 mm. The SPECT system sensitivity for a FOV of 20 mm and 15% energy window was 700 cps/MBq (7.8 × 10−2%) using a multi-pinhole equipped with five apertures 1 mm in diameter, whereas the PET absolute sensitivity was 2% for a 350–650 keV energy window and a 5 ns timing window. Several animal images are also presented.
Conclusions:
The new small animal PET/SPECT/CT proposed here exhibits high performance, producing high-quality images suitable for studies with small animals. Monolithic design for PET and SPECT scintillator crystals reduces cost and complexity without significant performance degradation.
Noble element time projection chambers are a leading technology for rare event detection in physics, such as for dark matter and neutrinoless double beta decay searches. Time projection chambers ...typically assign event position in the drift direction using the relative timing of prompt scintillation and delayed charge collection signals, allowing for reconstruction of an absolute position in the drift direction. In this paper, alternate methods for assigning event drift distance via quantification of electron diffusion in a pure high pressure xenon gas time projection chamber are explored. Data from the NEXT-White detector demonstrate the ability to achieve good position assignment accuracy for both high- and low-energy events. Using point-like energy deposits from
83
m
Kr calibration electron captures (
E
∼
45
keV), the position of origin of low-energy events is determined to 2 cm precision with bias
<
1
mm. A convolutional neural network approach is then used to quantify diffusion for longer tracks (
E
≥
1.5
MeV), from radiogenic electrons, yielding a precision of 3 cm on the event barycenter. The precision achieved with these methods indicates the feasibility energy calibrations of better than 1% FWHM at Q
β
β
in pure xenon, as well as the potential for event fiducialization in large future detectors using an alternate method that does not rely on primary scintillation.
A new method to tag the barium daughter in the double-beta decay of ^{136}Xe is reported. Using the technique of single molecule fluorescent imaging (SMFI), individual barium dication (Ba^{++}) ...resolution at a transparent scanning surface is demonstrated. A single-step photobleach confirms the single ion interpretation. Individual ions are localized with superresolution (∼2 nm), and detected with a statistical significance of 12.9σ over backgrounds. This lays the foundation for a new and potentially background-free neutrinoless double-beta decay technology, based on SMFI coupled to high pressure xenon gas time projection chambers.
A
bstract
The NEXT experiment aims at the sensitive search of the neutrinoless double beta decay in
136
Xe, using high-pressure gas electroluminescent time projection chambers. The NEXT-White ...detector is the first radiopure demonstrator of this technology, operated in the Laboratorio Subterráneo de Canfranc. Achieving an energy resolution of 1% FWHM at 2.6 MeV and further background rejection by means of the topology of the reconstructed tracks, NEXT-White has been exploited beyond its original goals in order to perform a neu- trinoless double beta decay search. The analysis considers the combination of 271.6 days of
136
Xe-enriched data and 208.9 days of
136
Xe-depleted data. A detailed background modeling and measurement has been developed, ensuring the time stability of the radiogenic and cosmogenic contributions across both data samples. Limits to the neutrinoless mode are obtained in two alternative analyses: a background-model-dependent approach and a novel direct background-subtraction technique, offering results with small dependence on the background model assumptions. With a fiducial mass of only 3.50 ± 0.01 kg of
136
Xe-enriched xenon, 90% C.L. lower limits to the neutrinoless double beta decay are found in the
T
1
/
2
0
ν
> 5
.
5
×
10
23
−
1
.
3
×
10
24
yr range, depending on the method. The presented techniques stand as a proof-of-concept for the searches to be implemented with larger NEXT detectors.
Abstract We suggest a symmetric-polar pixellation scheme which makes possible a reduction of the computational cost for expectation maximization (EM) iterative algorithms. The proposed ...symmetric-polar pixellation allows us to deal with 3D images as a whole problem without dividing the 3D problem into 2D slices approach. Performance evaluation of each approach in terms of stability and image quality is presented. Exhaustive comparisons between all approaches were conducted in a 2D based image reconstruction model. From these 2D approaches, that showing the best performances were finally implemented and evaluated in a 3D based image reconstruction model. Comparison to 3D images reconstructed with FBP is also presented. Although the algorithm is presented in the context of computed tomography (CT) image reconstruction, it can be applied to any other tomographic technique as well, due to the fact that the only requirement is a scanning geometry involving measurements of an object under different projection angles. Real data have been acquired with a small animal (CT) scanner to verify the proposed mathematical description of the CT system.
Purpose:
The authors have developed a small animal Positron emission tomography (PET) scanner based on monolithic LYSO crystals coupled to multi-anode photomultiplier tubes (MA-PMTs). In this study, ...the authors report on the design, calibration procedure, and performance evaluation of a PET system that the authors have developed using this innovative nonpixelated detector design.
Methods
: The scanner is made up of eight compact modules forming an octagon with an axial field of view (FOV) of 40 mm and a transaxial FOV of 80 mm diameter. In order to fully determine its performance, a recently issued National Electrical Manufacturers Association (NEMA) NU-4 protocol, specifically developed for small animal PET scanners, has been followed. By measuring the width of light distribution collected in the MA-PMT the authors are able to determine depth of interaction (DOI), thus making the proper identification of lines of response (LORs) with large incidence angles possible. PET performances are compared with those obtained with currently commercially available small animal PET scanners.
Results
: At axial center when the point-like source is located at 5 mm from the radial center, the spatial resolution measured was 1.65, 1.80, and 1.86 mm full width at half maximum (FWHM) for radial, tangential, and axial image profiles, respectively. A system scatter fraction of 7.5% (mouse-like phantom) and 13% (rat-like phantom) was obtained, while the maximum noise equivalent count rate (NECR) was 16.9 kcps at 12.7 MBq (0.37 MBq/ml) for mouse-like phantom and 12.8 kcps at 12.4 MBq (0.042 MBq/ml) for rat-like phantom The peak absolute sensitivity in the center of the FOV is 2% for a 30% peak energy window. Several animal images are also presented.
Conclusions:
The overall performance of our small animal PET is comparable to that obtained with much more complex crystal pixelated PET systems. Moreover, the new proposed PET produces high-quality images suitable for studies with small animals.