This study compares HS posturography on inertial sensors (MediPost) with force platform posturography in patients with unilateral vestibular dysfunction. The study group included 38 patients (age ...50.6; SD 11.6) with unilateral vestibular weakness (UV) and 65 healthy volunteers (48.7; SD 11.5). HS tests were performed simultaneously on the force plate and with MediPost sensor attached at L4. Four conditions applied: eyes open/closed, firm/foam. The tests were performed twice, with the head moving at the frequency of 0.3 Hz (HS 0.3) and 0.6 Hz (HS 0.6). Mean sway velocity was significantly lower for MediPost than force plate in 4th condition both in UV and healthy group. For HS 0.3 the differences between devices were marginal; the highest sensitivity (87%) and specificity (95%) were in 4th condition. For HS 0.6 MediPost revealed lower sensitivity than force plate although the surface parameter improved results. MediPost IMU device and force platform posturography revealed a similar ability to differentiate between patients with balance problems in course of vestibular pathology and healthy participants, despite the differences observed between measuring methods. In some tests surface parameter may be more appropriate than sway velocity in improving MediPost sensitivity.
The aim of our study was to validate the method of head-shake static posturography (HS-posturography) in healthy individuals and to establish the value of this novel method in the diagnostics of ...patients with unilateral vestibular lesion (UV). The study included 202 participants divided into two groups, one consisting of 133 patients with canal paresis CP > 19% and one of 69 healthy subjects. Participant was tested according to the standard protocol of static posturography (SP), and with head movements of 0.3 Hz (HS 40), 0.6 Hz (HS 70) in random order controlled by a metronome. HS-posturography revealed a similar repeatability and internal consistency as the standard posturography. In patients with UV, 4th condition revealed higher sensitivity (74%) and specificity (71%) in HS 40 than in the standard posturography (67%, 65% respectively) and HS 70 (54%, 70% respectively). Static posturography and HS- posturography revealed a high reliability of the testing method. The head movements added to static posturography improve the sensitivity and specificity of the method in group with vestibular impairment. The most important test for that purpose seems to be the one on unstable surface with the eyes closed, with low frequency of head movements.
(1) Objectives: The evaluation of dizzy patients is difficult due to nonspecific symptoms that require a multi-specialist approach. The Dizziness Handicap Inventory (DHI) is widely used in the ...assessment of dizziness-related disability, but its clinical efficacy needs further expansion. The aim of this study was to identify the subscales of DHI that may correlate with some vestibular or nonvestibular dysfunctions. (2) Material and methods: This observational study included 343 dizzy patients with one of the following clinical conditions: Vestibular impairment noncompensated or compensated, central or bilateral, benign paroxysmal positional vertigo (BPPV), migraine and psychogenic dizziness. Principal component analysis was used to examine the factorial structure of the questionnaire. (3) Results: The DHI questionnaire total scoring and its vestibular subscale distinguished between patients with compensated and uncompensated vestibular dysfunction with positive predictive values of 76% and 79%, respectively. The DHI items composing the F3 (positional) subscale revealed the highest scoring in the BPPV group with 75% sensitivity and 92% negative predictive value (NPV) in reference to Dix-Hallpike tests. The DHI total score and the subscales scores correlated with anxiety-depression, and the highest correlation coefficients were calculated for vestibular (F2 0.56) and anxiety (F5 0.51) subscales. (4) Conclusions: Our analysis revealed that the DHI vestibular subscale distinguishes between patients with compensated and uncompensated vestibular dysfunction. The positional subscale showed the highest scoring in the BPPV group with high sensitivity and low specificity of the test. The DHI is highly correlated with patients' psychological status.
This paper presents a fall risk assessment approach based on a fast mobility test, automatically evaluated using a low-cost, scalable system for the recording and analysis of body movement. This ...mobility test has never before been investigated as a sole source of data for fall risk assessment. It can be performed in a very limited space and needs only minimal additional equipment, yet provides large amounts of information, as the presented system can obtain much more data than traditional observation by capturing minute details regarding body movement. The readings are provided wirelessly by one to seven low-cost micro-electro-mechanical inertial measurement units attached to the subject's body segments. Combined with a body model, these allow segment rotations and translations to be computed and for body movements to be recreated in software. The subject can then be automatically classified by an artificial neural network based on selected values in the test, and those with an elevated risk of falls can be identified. Results obtained from a group of 40 subjects of various ages, both healthy volunteers and patients with vestibular system impairment, are presented to demonstrate the combined capabilities of the test and system. Labelling of subjects as fallers and non-fallers was performed using an objective and precise sensory organization test; it is an important novelty as this approach to subject labelling has never before been used in the design and evaluation of fall risk assessment systems. The findings show a true-positive ratio of 85% and true-negative ratio of 63% for classifying subjects as fallers or non-fallers using the introduced fast mobility test, which are noticeably better than those obtained for the long-established Timed Up and Go test.
The functions of most proteins are yet to be determined. The function of an enzyme is often defined by its interacting partners, including its substrate and product, and its role in larger metabolic ...networks. Here, we describe a computational method that predicts the functions of orphan enzymes by organizing them into a linear metabolic pathway. Given candidate enzyme and metabolite pathway members, this aim is achieved by finding those pathways that satisfy structural and network restraints implied by varied input information, including that from virtual screening, chemoinformatics, genomic context analysis, and ligand -binding experiments. We demonstrate this integrative pathway mapping method by predicting the L-gulonate catabolic pathway in
Rd KW20. The prediction was subsequently validated experimentally by enzymology, crystallography, and metabolomics. Integrative pathway mapping by satisfaction of structural and network restraints is extensible to molecular networks in general and thus formally bridges the gap between structural biology and systems biology.
Virtual reality (VR) is a well-established technology in medicine. Head-mounted displays (HMDs) have made VR more accessible in many branches of medical research. However, its application in balance ...evaluation has been vague, and comprehensive literature on possible applications of VR in posture measurement is scarce. The aim of this review is to conduct a literature search on the application of immersive VR delivered using a head-mounted display in posturographic measurements. A systematic search of two databases, PubMed and Scopus, using the keywords “virtual reality” and “posturography,” was performed following PRISMA guidelines for systematic reviews. Initial search results returned 89 non-duplicate records. Two reviewers independently screened the abstracts. Sixteen papers fulfilled the inclusion criteria and none of the exclusion criteria and were selected for complete text retrieval. An additional 16 records were identified from citation searching. Ultimately, 21 studies were included in this review. virtual reality is often used as additional visual stimuli in static and dynamic posturography evaluation. Only one study has attempted to evaluate a VR environment in a head-mounted display as an independent method in the assessment of posture. Further research should be conducted to assess HMD VR as a standalone posturography replacement.
This review summarizes current knowledge about the occurrence of hearing and balance disorders after antimalarial drugs treatment. It also examines the clinical applications of antimalarials, their ...mechanisms behind this ototoxicity and how it can be monitored. It includes studies with larger numbers of patients and those in which auditory function was assessed using audiological tests. Some antimalarials have been repurposed for other conditions like autoimmune disorders, rheumatic diseases, some viral diseases and cancers. While old antimalarial drugs, such as quinoline derivatives, are known to demonstrate ototoxicity, a number of new synthetic antimalarial agents particularly artemisinin derivatives, demonstrate unknown ototoxicity. Adverse audiovestibular effects vary depending on the medication itself, its dose and route of administration, as well as the drug combination, treated disease and individual predispositions of the patient. Dizziness was commonly reported, while vestibular symptoms, hearing loss and tinnitus were observed much less frequently, and most of these symptoms were reversible. As early identification of ototoxic hearing loss is critical to introducing possible alternative treatments with less ototoxic medications, therefore monitoring systems of those drugs ototoxic side effects are much needed.
Recently, two types of movement sensors have been introduced into Virtual Reality (VR) therapy: motion trackers and force-plate platforms. Combining these two methods could produce better ...rehabilitation outcomes. Such devices, encompassing motion trackers and force platforms, are referred to as "hybrid" VR units.
To assess the effectiveness of a low-cost hybrid VR based vestibular rehabilitation programMETHODS: A prospective, non-randomized, controlled group study comparing training using a hybrid VR unit (Group 1 n = 25) vs. static posturography with visual feedback (Group 2 n = 25) in patients with peripheral vestibular dysfunction was conducted. The subjects underwent 10 training sessions over 10 days (30 minute sessions). All were examined on a posturography platform at the start and 1 month after rehabilitation and completed the Vertigo Symptom Scale - Short Form (VSS-SF) questionnaire.
Both groups demonstrated improvement in posturographic parameters, which were statistically significant, but when comparing results between both groups there were no differences. The patients reported improvement in their subjective perception of symptoms on the VSS-SF scale, which were statistically significant in both groups, but greater in the VR group.
Both methods reduce postural sway, however subjective reduction of symptoms was greater in the VR group.