Although the accessibility of HIV treatment in developing nations has increased dramatically over the past decade, viral load testing to monitor the response of patients receiving therapy is often ...unavailable. Existing viral load technologies are often too expensive or resource-intensive for poor settings, and there is no appropriate HIV viral load test currently available at the point-of-care in low resource settings. Here, we present a lateral flow assay that employs gold nanoparticle probes and gold enhancement solution to detect amplified HIV RNA quantitatively. Preliminary results show that, when coupled with nucleic acid sequence based amplification (NASBA), this assay can detect concentrations of HIV RNA that match the clinically relevant range of viral loads found in HIV patients. The lateral flow test is inexpensive, simple and rapid to perform, and requires few resources. Our results suggest that the lateral flow assay may be integrated with amplification and sample preparation technologies to serve as an HIV viral load test for low-resource settings.
Despite the importance of early diagnosis and treatment of HIV, only a small fraction of HIV-exposed infants in low- and middle-income countries are tested for the disease. The gold standard for ...early infant diagnosis, DNA PCR, requires resources that are unavailable in poor settings, and no point-of-care HIV DNA test is currently available. We have developed a device constructed of layers of paper, glass fiber, and plastic that is capable of performing isothermal, enzymatic amplification of HIV DNA. The device is inexpensive, small, light-weight, and easy to assemble. The device stores lyophilized enzymes, facilitates mixing of reaction components, and supports recombinase polymerase amplification in five steps of operation. Using commercially available lateral flow strips as a detection method, we demonstrate the ability of our device to amplify 10 copies of HIV DNA to detectable levels in 15 min. Our results suggest that our device, which is designed to be used after DNA extraction from dried-blood spots, may serve in conjunction with lateral flow strips as part of a point-of-care HIV DNA test to be used in low resource settings.
Blood obtained via fingerprick is commonly used in point-of-care assays, but few studies have assessed variability in parameters obtained from successive drops of fingerprick blood, which may cause ...problems for clinical decision making and for assessing accuracy of point-of-care tests.
We used a hematology analyzer to analyze the hemoglobin concentration, total WBC count, three-part WBC differential, and platelet count in six successive drops of blood collected from one fingerprick from each of 11 donors, and we used a hemoglobinometer to measure the hemoglobin concentration of 10 drops of fingerprick blood from each of 7 donors.
The average percent coefficient of variation (CV) for successive drops of fingerprick blood was higher by up to 3.4 times for hemoglobin, 5.7 times for WBC count, 3 times for lymphocyte count, 7.7 times for granulocyte count, and 4 times for platelets than in venous controls measured using a hematology analyzer. The average percent CV for fingerprick blood was up to 5 times higher for hemoglobin than venous blood measured using a point-of-care hemoglobinometer. Fluctuations in blood parameters with increasing volume of fingerprick blood are within instrument variability for volumes equal to or greater than 60 to 100 μL.
These data suggest caution when using measurements from a single drop of fingerprick blood.
Respiratory failure is a leading cause of neonatal mortality in the developing world. Bubble continuous positive airway pressure (bCPAP) is a safe, effective intervention for infants with respiratory ...distress and is widely used in developed countries. Because of its high cost, bCPAP is not widely utilized in low-resource settings. We evaluated the performance of a new bCPAP system to treat severe respiratory distress in a low resource setting, comparing it to nasal oxygen therapy, the current standard of care.
We conducted a non-randomized convenience sample study to test the efficacy of a low-cost bCPAP system treating newborns with severe respiratory distress in the neonatal ward of Queen Elizabeth Central Hospital, in Blantyre, Malawi. Neonates weighing >1,000 g and presenting with severe respiratory distress who fulfilled inclusion criteria received nasal bCPAP if a device was available; if not, they received standard care. Clinical assessments were made during treatment and outcomes compared for the two groups.
87 neonates (62 bCPAP, 25 controls) were recruited. Survival rate for neonates receiving bCPAP was 71.0% (44/62) compared with 44.0% (11/25) for controls. 65.5% (19/29) of very low birth weight neonates receiving bCPAP survived to discharge compared to 15.4% (1/13) of controls. 64.6% (31/48) of neonates with respiratory distress syndrome (RDS) receiving bCPAP survived to discharge, compared to 23.5% (4/17) of controls. 61.5% (16/26) of neonates with sepsis receiving bCPAP survived to discharge, while none of the seven neonates with sepsis in the control group survived.
Use of a low-cost bCPAP system to treat neonatal respiratory distress resulted in 27% absolute improvement in survival. The beneficial effect was greater for neonates with very low birth weight, RDS, or sepsis. Implementing appropriate bCPAP devices could reduce neonatal mortality in developing countries.
Cervical cancer remains a leading cause of cancer death for women in low‐ and middle‐income countries. The goal of our study was to evaluate screening and triage strategies, including high‐resolution ...microendoscopy (HRME), to detect cervical abnormalities concerning for precancer at the point of care. Women (n = 1824) were enrolled at the Instituto de Cáncer de El Salvador. All underwent screening by both human papillomavirus (HPV) testing using careHPV and visual inspection with acetic acid (VIA). Screen‐positives, along with 10% of screen‐negatives, were invited to return for a follow‐up examination that included triage with VIA, colposcopy and HRME imaging. Biopsies were taken of any abnormalities identified. If no abnormalities were identified, then the worst scoring site by HRME was biopsied. The sensitivities of HPV testing and VIA to screen for cervical intraepithelial neoplasia Grade 2 or more severe diagnoses (CIN2+) were 82.1% and 75% (P = .77), while the specificities were 90.4% and 80.9% (P < .001), respectively. The sensitivities of VIA, colposcopy and HRME as triage tests for CIN2+ were 82.1%, 82.1% and 71.4%, respectively (P ≥ .38). HRME had a significantly higher specificity (66.7%) than VIA (51.9%) (P < .001) and colposcopy (53.3%) (P < .001). When evaluating different theoretical screening and triage strategies, screening with HPV testing followed by triage with HRME would result in more women receiving appropriate care (97%) compared to screening with VIA (75%) or HPV alone (90%). Our findings demonstrate that screening with HPV is superior to VIA, and that triage with HRME imaging increases the specificity of detecting CIN2+ at the point of care in a low‐resource setting.
What's new?
Triage with a new high‐resolution microscopy technique improved the specificity of cervical screening in El Salvador. Visual inspection with acetic acid (VIA) is a low‐cost screening technique that requires no follow up visit. However, techniques with higher sensitivity and specificity would be preferable. High‐resolution microendoscopy (HRME) is a low‐cost fiber‐optic imaging device that allows providers to visualize the nuclei of cervical epithelial cells in vivo. Here, the authors compared five different screening and triage strategies. They demonstrated that screening with HPV testing is superior to VIA, and triage with HRME increased the specificity of detecting cervical precancer.
Potentially premalignant oral epithelial lesions (PPOELs) are a group of clinically suspicious conditions, of which a small percentage will undergo malignant transformation. PPOELs are suboptimally ...diagnosed and managed under the current standard of care. Dysplasia is the most well-established marker to distinguish high-risk PPOELs from low-risk PPOELs, and performing a biopsy to establish dysplasia is the diagnostic gold standard. However, a biopsy is limited by morbidity, resource requirements, and the potential for underdiagnosis. Diagnostic adjuncts may help clinicians better evaluate PPOELs before definitive biopsy, but existing adjuncts, such as toluidine blue, acetowhitening, and autofluorescence imaging, have poor accuracy and are not generally recommended. Recently, in vivo microscopy technologies, such as high-resolution microendoscopy, optical coherence tomography, reflectance confocal microscopy, and multiphoton imaging, have shown promise for improving PPOEL patient care. These technologies allow clinicians to visualize many of the same microscopic features used for histopathologic assessment at the point of care.
Background
Multimodal optical imaging, incorporating reflectance and fluorescence modalities, is a promising tool to detect oral premalignant lesions in real‐time.
Methods
Images were acquired from ...171 sites in 66 patient visits for clinical evaluation of oral lesions. An automated algorithm was used to classify lesions as high‐ or low‐risk for neoplasia. Biopsies were acquired at clinically indicated sites and those classified as high‐risk by imaging, at the surgeon's discretion.
Results
Twenty sites were biopsied based on clinical examination or imaging. Of these, 12 were indicated clinically and by imaging; 58% were moderate dysplasia or worse. Four biopsies were indicated by imaging evaluation only; 75% were moderate dysplasia or worse. Finally, four biopsies were indicated by clinical evaluation only; 75% were moderate dysplasia or worse.
Conclusion
Multimodal imaging identified more cases of high‐grade dysplasia than clinical evaluation, and can improve detection of high grade precancer in patients with oral lesions.
Histopathology plays a critical role in the diagnosis and surgical management of cancer. However, access to histopathology services, especially frozen section pathology during surgery, is limited in ...resource-constrained settings because preparing slides from resected tissue is time-consuming, labor-intensive, and requires expensive infrastructure. Here, we report a deep-learning-enabled microscope, named DeepDOF-SE, to rapidly scan intact tissue at cellular resolution without the need for physical sectioning. Three key features jointly make DeepDOF-SE practical. First, tissue specimens are stained directly with inexpensive vital fluorescent dyes and optically sectioned with ultra-violet excitation that localizes fluorescent emission to a thin surface layer. Second, a deep-learning algorithm extends the depth-of-field, allowing rapid acquisition of in-focus images from large areas of tissue even when the tissue surface is highly irregular. Finally, a semi-supervised generative adversarial network virtually stains DeepDOF-SE fluorescence images with hematoxylin-and-eosin appearance, facilitating image interpretation by pathologists without significant additional training. We developed the DeepDOF-SE platform using a data-driven approach and validated its performance by imaging surgical resections of suspected oral tumors. Our results show that DeepDOF-SE provides histological information of diagnostic importance, offering a rapid and affordable slide-free histology platform for intraoperative tumor margin assessment and in low-resource settings.