  • A Simple Low-Dose X-Ray CT ...
    Zeng, Dong; Huang, Jing; Bian, Zhaoying; Niu, Shanzhou; Zhang, Hua; Feng, Qianjin; Liang, Zhengrong; Ma, Jianhua

    IEEE transactions on nuclear science, 10/2015, Letnik: 62, Številka: 5
    Journal Article

    Low-dose X-ray computed tomography (CT) simulation from a high-dose scan is required in optimizing radiation dose to patients. In this paper, we propose a simple low-dose CT simulation strategy in the sinogram domain using the raw data from high-dose scan. Specially, a relationship between the incident fluxes of low- and high-dose scans is first determined according to the repeated projection measurements and analysis. Second, the incident flux level of the simulated low-dose scan is generated by properly scaling the incident flux level of the high-dose scan via the determined relationship in the first step. Third, the low-dose CT transmission data by energy integrating detection is simulated by adding a statistically independent Poisson noise distribution plus a statistically independent Gaussian noise distribution. Finally, a filtered back-projection (FBP) algorithm is implemented to reconstruct the resultant low-dose CT images. The present low-dose simulation strategy is verified on the simulations and real scans by comparing it with the existing low-dose CT simulation tool. Experimental results demonstrated that the present low-dose CT simulation strategy can generate accurate low-dose CT sinogram data from high-dose scans in terms of qualitative and quantitative measurements.