Abstract
The radioiodine isotope pair
124
I/
131
I is used in a theranostic approach for patient-specific treatment of differentiated thyroid cancer. Lesion detectability is notably higher for
124
I ...PET (positron emission tomography) than for
131
I gamma camera imaging but can be limited for small and low uptake lesions. The recently introduced silicon-photomultiplier-based (SiPM-based) PET/CT (computed tomography) systems outperform previous-generation systems in detector sensitivity, coincidence time resolution, and spatial resolution. Hence, SiPM-based PET/CT shows an improved detectability, particularly for small lesions. In this study, we compare the size-dependant minimum detectable
124
I activity (MDA) between the SiPM-based Biograph Vision and the previous-generation Biograph mCT PET/CT systems and we attempt to predict the response to
131
I radioiodine therapy of lesions additionally identified on the SiPM-based system. A tumour phantom mimicking challenging conditions (derived from published patient data) was used; i.e., 6 small spheres (diameter of 3.7–9.7 mm), 9 low activity concentrations (0.25–25 kBq/mL), and a very low signal-to-background ratio (20:1). List-mode emission data (single-bed position) were divided into frames of 4, 8, 16, and 30 min. Images were reconstructed with ordinary Poisson ordered-subsets expectation maximization (OSEM), additional time-of-flight (OSEM-TOF) or TOF and point spread function modelling (OSEM-TOF+PSF). The signal-to-noise ratio and the MDA were determined. Absorbed dose estimations were performed to assess possible treatment response to high-activity
131
I radioiodine therapy. The signal-to-noise ratio and the MDA were improved from the mCT to the Vision, from OSEM to OSEM-TOF and from OSEM-TOF to OSEM-TOF+PSF reconstructed images, and from shorter to longer emission times. The overall mean MDA ratio of the Vision to the mCT was 0.52 ± 0.18. The absorbed dose estimations indicate that lesions ≥ 6.5 mm with expected response to radioiodine therapy would be detectable on both systems at 4-min emission time. Additional smaller lesions of therapeutic relevance could be detected when using a SiPM-based PET system at clinically reasonable emission times. This study demonstrates that additional lesions with predicted response to
131
I radioiodine therapy can be detected. Further clinical evaluation is warranted to evaluate if negative
124
I PET scans on a SiPM-based system can be sufficient to preclude patients from blind radioiodine therapy.
Background New-generation silicon-photomultiplier (SiPM)-based PET/CT systems exhibit an improved lesion detectability and image quality due to a higher detector sensitivity. Consequently, the ...acquisition time can be reduced while maintaining diagnostic quality. The aim of this study was to determine the lowest .sup.18F-FDG PET acquisition time without loss of diagnostic information and to optimise image reconstruction parameters (image reconstruction algorithm, number of iterations, voxel size, Gaussian filter) by phantom imaging. Moreover, patient data are evaluated to confirm the phantom results. Methods Three phantoms were used: a soft-tissue tumour phantom, a bone-lung tumour phantom, and a resolution phantom. Phantom conditions (lesion sizes from 6.5 mm to 28.8 mm in diameter, lesion activity concentration of 15 kBq/mL, and signal-to-background ratio of 5:1) were derived from patient data. PET data were acquired on an SiPM-based Biograph Vision PET/CT system for 10 min in list-mode format and resampled into time frames from 30 to 300 s in 30-s increments to simulate different acquisition times. Different image reconstructions with varying iterations, voxel sizes, and Gaussian filters were probed. Contrast-to-noise-ratio (CNR), maximum, and peak signal were evaluated using the 10-min acquisition time image as reference. A threshold CNR value greater than or equal to 5 and a maximum (peak) deviation of + or - 20% were considered acceptable. 20 patient data sets were evaluated regarding lesion quantification as well as agreement and correlation between reduced and full acquisition time standard uptake values (assessed by Pearson correlation coefficient, intraclass correlation coefficient, Bland-Altman analyses, and Krippendorff's alpha). Results An acquisition time of 60 s per bed position yielded acceptable detectability and quantification results for clinically relevant phantom lesions greater than or equal to 9.7 mm in diameter using OSEM-TOF or OSEM-TOF+PSF image reconstruction, a 4-mm Gaussian filter, and a 1.65 x 1.65 x 2.00-mm.sup.3 or 3.30 x 3.30 x 3.00-mm.sup.3 voxel size. Correlation and agreement of patient lesion quantification between full and reduced acquisition times were excellent. Conclusion A threefold reduction in acquisition time is possible. Patients might benefit from more comfortable examinations or reduced radiation exposure, if instead of the acquisition time the applied activity is reduced. Keywords: Positron emission tomography, FDG, Acquisition time, Silicon-based photomultiplier, Digital PET, Protocol optimisation, Lymphoma
To develop and evaluate a new approach for spatially variant and tissue-dependent positron range (PR) correction (PRC) during the iterative PET image reconstruction.
The PR distributions of three ...radionuclides (
F,
Ga, and
I) were simulated using the GATE (GEANT4) framework in different material compositions (lung, water, and bone). For every radionuclide, the uniform PR kernel was created by mapping the simulated 3D PR point cloud to a 3D matrix with its size defined by the maximum PR in lung (
F) or water (
Ga and
I) and the PET voxel size. The spatially variant kernels were composed from the uniform PR kernels by analyzing the material composition of the surrounding medium for each voxel before implementation as tissue-dependent, point-spread functions into the iterative image reconstruction. The proposed PRC method was evaluated using the NEMA image quality phantom (
F,
Ga, and
I); two unique PR phantoms were scanned and evaluated following OSEM reconstruction with and without PRC using different metrics, such as contrast recovery, contrast-to-noise ratio, image noise and the resolution evaluated in terms of full width at half maximum (FWHM).
The effect of PRC on
F-imaging was negligible. In contrast, PRC improved image contrast for the 10-mm sphere of the NEMA image quality phantom filled with
Ga and
I by 33 and 24%, respectively. While the effect of PRC was less noticeable for the larger spheres, contrast recovery still improved by 5%. The spatial resolution was improved by 26% for
I (FWHM of 4.9 vs. 3.7 mm).
For high energy positron-emitting radionuclides, the proposed PRC method helped recover image contrast with reduced noise levels and with improved spatial resolution. As such, the PRC approach proposed here can help improve the quality of PET data in clinical practice and research.
Transthyretin (ATTR) amyloidosis is responsible for the majority of cardiac amyloidosis (CA) cases and can be reliably diagnosed with bone scintigraphy and the visual Perugini score. We aimed to ...implement a quantification method of cardiac amyloid deposits in patients with suspected cardiac amyloidosis and to compare performance to visual scoring.
136 patients received 99mTc-DPD-bone scintigraphy including SPECT/CT of the thorax in case of suspicion of cardiac amyloidosis. Imaging phantom studies were performed to determine the scaling factor for standardized uptake value (SUV) quantification from SPECT/CT. Myocardial tracer uptake was quantified in a whole heart volume of interest.
Forty-five patients were diagnosed with CA. A strong relationship between cardiac SUVmax and Perugini score was found (Spearman r 0.75, p < 0.0001). Additionally, tracer uptake in bone decreased with increasing cardiac SUVmax and Perugini score (p < 0.0001). ROC analysis revealed good performance of the SUVmax for the detection of ATTR-CA with AUC of 0.96 ± 0.02 (p < 0.0001) with sensitivity 98.7% and specificity 87.2%.
We demonstrate an accessible and accurate quantitative SPECT approach in CA. Quantitative assessment of the cardiac tracer uptake may improve diagnostic accuracy and risk classification. This method may enable monitoring and assessment of therapy response in patients with ATTR amyloidosis.
Background The superior accuracy and sensitivity of .sup.18F-FDG-PET/CT in comparison to morphological imaging alone leads to an upstaging in up to 30% of lymphoma patients. Novel digital PET/CT ...scanners might enable to reduce administered tracer activity or scan time duration while maintaining diagnostic performance; this might allow for a higher patient throughput or a reduced radiation exposure, respectively. In particular, the radiation exposure reduction is of interest due to the often young age and high remission rate of lymphoma patients. Methods Twenty patients with (suspected) lymphoma (6 for initial staging, 12 after systemic treatment, 2 in suspicion of recurrence) sequentially underwent .sup.18F-FDG-PET/CT examinations on a digital PET/CT (Siemens Biograph Vision) with a total scan time duration of 15 min (reference acquisition protocol) and 5 min (reduced acquisition protocol) using continuous-bed-motion. Both data sets were reconstructed using either standalone time of flight (TOF) or in combination with point spread function (PSF), each with 2 and 4 iterations. Lesion detectability by blinded assessment (separately for supra- and infradiaphragmal nodal lesions and for extranodal lesions), lesion image quantification, and image noise were used as metrics to assess diagnostic performance. Additionally, Deauville Score was compared for all patients after systemic treatment. Results All defined regions were correctly classified in the images acquired with reduced emission time, and therefore, no changes in staging were observed. Lesion quantification was acceptable, that is, mean absolute percentage deviation of maximum and peak standardized uptake values were 6.8 and 6.4% (derived from 30 lesions). A threefold reduction of scan time duration led to an increase in image noise from 7.1 to 11.0% (images reconstructed with 4 iterations) and from 4.7 to 7.2% (images reconstructed with 2 iterations). No deviations in Deauville Score were observed. Conclusion These results suggest that scan time duration or administered tracer activity can be reduced threefold without compromising diagnostic performance. Especially a reduction of administered activity might allow for a lower radiation exposure and better health economics. Larger trials are warranted to confirm our results. Keywords: PET/CT, FDG, Image quality, Silicon photomultiplier, Lymphoma
Aim
68
GaGa-PSMA-11 PET/CT allows for a superior detection of prostate cancer tissue, especially in the context of a low tumor burden. Digital PET/CT bears the potential of reducing scan time ...duration/administered tracer activity due to, for instance, its higher sensitivity and improved time coincidence resolution. It might thereby expand
68
GaGa-PSMA-11 PET/CT that is currently limited by
68
Ge/
68
Ga-generator yield. Our aim was to clinically evaluate the influence of a reduced scan time duration in combination with different image reconstruction algorithms on the diagnostic performance.
Methods
Twenty prostate cancer patients (11 for biochemical recurrence, 5 for initial staging, 4 for metastatic disease) sequentially underwent
68
GaGa-PSMA-11 PET/CT on a digital Siemens Biograph Vision. PET data were collected in continuous-bed-motion mode with a mean scan time duration of 16.7 min (reference acquisition protocol) and 4.6 min (reduced acquisition protocol). Four iterative reconstruction algorithms were applied using a time-of-flight (TOF) approach alone or combined with point-spread-function (PSF) correction, each with 2 or 4 iterations. To evaluate the diagnostic performance, the following metrics were chosen: (a) per-region detectability, (b) the tumor maximum and peak standardized uptake values (SUVmax and SUVpeak), and (c) image noise using the liver’s activity distribution.
Results
Overall, 98% of regions (91% of affected regions) were correctly classified in the reduced acquisition protocol independent of the image reconstruction algorithm. Two nodal lesions (each ≤ 4 mm) were not identified (leading to downstaging in 1/20 cases). Mean absolute percentage deviation of SUVmax (SUVpeak) was approximately 9% (6%) for each reconstruction algorithm. The mean image noise increased from 13 to 21% (4 iterations) and from 10 to 15% (2 iterations) for PSF + TOF and TOF images.
Conclusions
High agreement at 3.5-fold reduction of scan time in terms of per-region detection (98% of regions) and image quantification (mean deviation ≤ 10%) was demonstrated; however, small lesions can be missed in about 10% of patients leading to downstaging (T1N0M0 instead of T1N1M0) in 5% of patients. Our results suggest that a reduction of scan time duration or administered
68
GaGa-PSMA-11 activities can be considered in metastatic patients, where missing small lesions would not impact patient management. Limitations include the small and heterogeneous sample size and the lack of follow-up.
Aim
Positron emission tomography (PET) using
124
I-mIBG has been established for imaging and pretherapeutic dosimetry. Here, we report the first systematic analysis of the biodistribution and ...radiation dosimetry of
124
I-mIBG in patients with neural crest tumours and project the results to paediatric patient models.
Methods
Adult patients with neural crest tumours who underwent sequential
124
I-mIBG PET were included in this retrospective single-center analysis. PET data were acquired 4, 24, 48, and/or 120 h after administration of a mean of 43 MBq
124
I-mIBG. Whole-body counting and blood sampling were performed at 2, 4, 24, 48 and 120 h after administration. Absorbed organ dose and effective dose coefficients were estimated in OLINDA/EXM 2.2 according to the MIRD formalism. Extrapolation to paediatric models was performed based on mass-fraction scaling of the organ-specific residence times. Biodistribution data for adults were also projected to
123
I-mIBG and
131
I-mIBG.
Results
Twenty-one patients (11 females, 10 males) were evaluated. For adults, the organs exposed to the highest dose per unit administered activity were urinary bladder (1.54 ± 0.40 mGy/MBq), salivary glands (0.77 ± 0.28 mGy/MBq) and liver (0.65 ± 0.22 mGy/MBq). Mean effective dose coefficient for adults was 0.25 ± 0.04 mSv/MBq (male: 0.24 ± 0.03 mSv/MBq, female: 0.26 ± 0.06 mSv/MBq), and increased gradually to 0.29, 0.44, 0.69, 1.21, and 2.94 mSv/MBq for the 15-, 10-, 5-, 1-years-old, and newborn paediatric reference patients. Projected mean effective dose coefficients for
123
I-mIBG and
131
I-mIBG for adults were 0.014 ± 0.002 mSv/MBq and 0.18 ± 0.04 mSv/MBq, respectively.
Conclusion
PET-based derived radiation dosimetry data for
124
I-mIBG from this study agreed well with historical projected data from ICRP 53. The effective dose coefficients presented here may aid in guidance for establishing weight-based activity administration protocols.
The use of recovery coefficients (RCs) in (124)I PET lesion imaging is a simple method to correct the imaged activity concentration (AC) primarily for the partial-volume effect and, to a minor ...extent, for the prompt gamma coincidence effect. The aim of this phantom study was to experimentally investigate a number of various factors affecting the (124)I RCs. Three RC-based correction approaches were considered. These approaches differ with respect to the volume of interest (VOI) drawn, which determines the imaged AC and the RCs: a single voxel VOI containing the maximum value (maximum RC), a spherical VOI with a diameter of the scanner resolution (resolution RC) and a VOI equaling the physical object volume (isovolume RC). Measurements were performed using mainly a stand-alone PET scanner (EXACT HR(+)) and a latest-generation PET/CT scanner (BIOGRAPH mCT). The RCs were determined using a cylindrical phantom containing spheres or rotational ellipsoids and were derived from images acquired with a reference acquisition protocol. For each type of RC, the influence of the following factors on the RC was assessed: object shape, background activity spill in and iterative image reconstruction parameters. To evaluate the robustness of the RC-based correction approaches, the percentage deviation between RC-corrected and true ACs was determined from images acquired with a clinical acquisition protocol of different AC regimes. The observed results of the shape and spill-in effects were compared with simulation data derived from a convolution-based model. The study demonstrated that the shape effect was negligible and, therefore, was in agreement with theoretical expectations. In contradiction to the simulation results, the observed spill-in effect was unexpectedly small. To avoid variations in the determination of RCs due to reconstruction parameter changes, image reconstruction with a pixel length of about one-third or less of the scanner resolution and an OSEM 1 x 32 algorithm or one with somewhat higher number of effective iterations are recommended. Using the clinical acquisition protocol, the phantom study indicated that the resolution- or isovolume-based recovery-correction approaches appeared to be more appropriate to recover the ACs from patient data; however, the application of the three RC-based correction approaches to small lesions containing low ACs was, in particular, associated with large underestimations. The phantom study had several limitations, which were discussed in detail.
Aim
To evaluate the effect of combining positron range correction (PRC) with point-spread-function (PSF) correction and to compare different methods of implementation into iterative image ...reconstruction for
124
I-PET imaging.
Materials and methods
Uniform PR blurring kernels of
124
I were generated using the GATE (GEANT4) framework in various material environments (lung, water, and bone) and matched to a 3D matrix. The kernels size was set to 11 × 11 × 11 based on the maximum PR in water and the voxel size of the PET system. PET image reconstruction was performed using the standard OSEM algorithm, OSEM with PRC implemented before the forward projection (OSEM+PRC simplified) and OSEM with PRC implemented in both forward- and back-projection steps (full implementation) (OSEM+PRC). Reconstructions were repeated with resolution recovery, point-spread function (PSF) included. The methods and kernel variation were validated using different phantoms filled with
124
I acquired on a Siemens mCT PET/CT system. The data was evaluated for contrast recovery and image noise.
Results
Contrast recovery improved by 2–10% and 4–37% with OSEM+PRC simplified and OSEM+PRC, respectively, depending on the sphere size of the NEMA IQ phantom. Including PSF in the reconstructions further improved contrast by 4–19% and 3–16% with the PSF+PRC simplified and PSF+PRC, respectively. The benefit of PRC was more pronounced within low-density material. OSEM-PRC and OSEM-PSF as well as OSEM-PSF+PRC in its full- and simplified implementation showed comparable noise and convergence. OSEM-PRC simplified showed comparably faster convergence but at the cost of increased image noise.
Conclusions
The combination of the PSF and PRC leads to increased contrast recovery with reduced image noise compared to stand-alone PSF or PRC reconstruction. For OSEM-PRC reconstructions, a full implementation in the reconstruction is necessary to handle image noise. For the combination of PRC with PSF, a simplified PRC implementation can be used to reduce reconstruction times.
Background
In recurrent differentiated thyroid cancer patients, detectability in
124
I PET is limited for lesions with low radioiodine uptake. We assess the improvements in lesion detectability and ...image quality between three generations of PET scanners with different detector technologies. The results are used to suggest an optimized protocol.
Methods
Datasets of 10 patients with low increasing thyroglobulin or thyroglobulin antibody levels after total thyroidectomy and radioiodine therapies were included. PET data were acquired and reconstructed on a Biograph mCT PET/CT (whole-body, 4-min acquisition time per bed position; OSEM, OSEM-TOF, OSEM-TOF+PSF), a non-TOF Biograph mMR PET/MR (neck region, 4 min and 20 min; OSEM), and a new generation Biograph Vision PET/CT (whole-body, 4 min; OSEM, OSEM-TOF, OSEM-TOF+PSF). The 20-min image on the mMR was used as reference to calculate the detection efficacy in the neck region. Image quality was rated on a 5-point scale.
Results
All detected lesions were in the neck region. Detection efficacy was 8/9 (Vision OSEM-TOF and OSEM-TOF+PSF), 4/9 (Vision OSEM), 3/9 (mMR OSEM and mCT OSEM-TOF+PSF), and 2/9 (mCT OSEM and OSEM-TOF). Median image quality was 4 (Vision OSEM-TOF and OSEM-TOF+PSF), 3 (Vision OSEM, mCT OSEM-TOF+PSF, and mMR OSEM 20-min), 2 (mCT OSEM-TOF), 1.5 (mCT OSEM), and 1 (mMR OSEM 4 min).
Conclusion
At a clinical standard acquisition time of 4 min per bed position, the new generation Biograph Vision using a TOF-based image reconstruction demonstrated the highest detectability and image quality and should, if available, be preferably used for imaging of low-uptake lesions. A prolonged acquisition time for the mostly affected neck region can be useful.