The feasibility, safety, and efficacy of prolonged use of an artificial beta cell (closed-loop insulin-delivery system) in the home setting have not been established.
In two multicenter, crossover, ...randomized, controlled studies conducted under free-living home conditions, we compared closed-loop insulin delivery with sensor-augmented pump therapy in 58 patients with type 1 diabetes. The closed-loop system was used day and night by 33 adults and overnight by 25 children and adolescents. Participants used the closed-loop system for a 12-week period and sensor-augmented pump therapy (control) for a similar period. The primary end point was the proportion of time that the glucose level was between 70 mg and 180 mg per deciliter for adults and between 70 mg and 145 mg per deciliter for children and adolescents.
Among adults, the proportion of time that the glucose level was in the target range was 11.0 percentage points (95% confidence interval CI, 8.1 to 13.8) greater with the use of the closed-loop system day and night than with control therapy (P<0.001). The mean glucose level was lower during the closed-loop phase than during the control phase (difference, -11 mg per deciliter; 95% CI, -17 to -6; P<0.001), as were the area under the curve for the period when the glucose level was less than 63 mg per deciliter (39% lower; 95% CI, 24 to 51; P<0.001) and the mean glycated hemoglobin level (difference, -0.3%; 95% CI, -0.5 to -0.1; P=0.002). Among children and adolescents, the proportion of time with the nighttime glucose level in the target range was higher during the closed-loop phase than during the control phase (by 24.7 percentage points; 95% CI, 20.6 to 28.7; P<0.001), and the mean nighttime glucose level was lower (difference, -29 mg per deciliter; 95% CI, -39 to -20; P<0.001). The area under the curve for the period in which the day-and-night glucose levels were less than 63 mg per deciliter was lower by 42% (95% CI, 4 to 65; P=0.03). Three severe hypoglycemic episodes occurred during the closed-loop phase when the closed-loop system was not in use.
Among patients with type 1 diabetes, 12-week use of a closed-loop system, as compared with sensor-augmented pump therapy, improved glucose control, reduced hypoglycemia, and, in adults, resulted in a lower glycated hemoglobin level. (Funded by the JDRF and others; AP@home04 and APCam08 ClinicalTrials.gov numbers, NCT01961622 and NCT01778348.).
The possible advantage of hybrid closed-loop therapy (i.e., artificial pancreas) over sensor-augmented pump therapy in very young children with type 1 diabetes is unclear.
In this multicenter, ...randomized, crossover trial, we recruited children 1 to 7 years of age with type 1 diabetes who were receiving insulin-pump therapy at seven centers across Austria, Germany, Luxembourg, and the United Kingdom. Participants received treatment in two 16-week periods, in random order, in which the closed-loop system was compared with sensor-augmented pump therapy (control). The primary end point was the between-treatment difference in the percentage of time that the sensor glucose measurement was in the target range (70 to 180 mg per deciliter) during each 16-week period. The analysis was conducted according to the intention-to-treat principle. Key secondary end points included the percentage of time spent in a hyperglycemic state (glucose level, >180 mg per deciliter), the glycated hemoglobin level, the mean sensor glucose level, and the percentage of time spent in a hypoglycemic state (glucose level, <70 mg per deciliter). Safety was assessed.
A total of 74 participants underwent randomization. The mean (±SD) age of the participants was 5.6±1.6 years, and the baseline glycated hemoglobin level was 7.3±0.7%. The percentage of time with the glucose level in the target range was 8.7 percentage points (95% confidence interval CI, 7.4 to 9.9) higher during the closed-loop period than during the control period (P<0.001). The mean adjusted difference (closed-loop minus control) in the percentage of time spent in a hyperglycemic state was -8.5 percentage points (95% CI, -9.9 to -7.1), the difference in the glycated hemoglobin level was -0.4 percentage points (95% CI, -0.5 to -0.3), and the difference in the mean sensor glucose level was -12.3 mg per deciliter (95% CI, -14.8 to -9.8) (P<0.001 for all comparisons). The time spent in a hypoglycemic state was similar with the two treatments (P = 0.74). The median time spent in the closed-loop mode was 95% (interquartile range, 92 to 97) over the 16-week closed-loop period. One serious adverse event of severe hypoglycemia occurred during the closed-loop period. One serious adverse event that was deemed to be unrelated to treatment occurred.
A hybrid closed-loop system significantly improved glycemic control in very young children with type 1 diabetes, without increasing the time spent in hypoglycemia. (Funded by the European Commission and others; ClinicalTrials.gov number, NCT03784027.).
Sarcopenia, a condition of low muscle mass, quality, and strength, is commonly found in patients with cirrhosis and is associated with adverse clinical outcomes including reduction in quality of ...life, increased mortality, and posttransplant complications. In chronic liver disease (CLD), sarcopenia is most commonly defined through the measurement of the skeletal muscle index of the third lumbar spine. A major contributor to sarcopenia in CLD is the imbalance in muscle protein turnover, which likely occurs due to a decrease in muscle protein synthesis and an elevation in muscle protein breakdown. This imbalance is assumed to arise due to several factors including accelerated starvation, hyperammonemia, amino acid deprivation, chronic inflammation, excessive alcohol intake, and physical inactivity. In particular, hyperammonemia is a key mediator of the liver-gut axis and is known to contribute to mitochondrial dysfunction and an increase in myostatin expression. Currently, the use of nutritional interventions such as late-evening snacks, branched-chain amino acid supplementation, and physical activity have been proposed to help the management and treatment of sarcopenia. However, little evidence exists to comprehensively support their use in clinical settings. Several new pharmacological strategies, including myostatin inhibition and the nutraceutical Urolithin A, have recently been proposed to treat age-related sarcopenia and may also be of use in CLD. This review highlights the potential molecular mechanisms contributing to sarcopenia in CLD alongside a discussion of existing and potential new treatment strategies.
The achievement of glycaemic control remains challenging for patients with type 1 diabetes. We assessed the effectiveness of day-and-night hybrid closed-loop insulin delivery compared with ...sensor-augmented pump therapy in people with suboptimally controlled type 1 diabetes aged 6 years and older.
In this open-label, multicentre, multinational, single-period, parallel randomised controlled trial, participants were recruited from diabetes outpatient clinics at four hospitals in the UK and two centres in the USA. We randomly assigned participants with type 1 diabetes aged 6 years and older treated with insulin pump and with suboptimal glycaemic control (glycated haemoglobin HbA1c 7·5–10·0%) to receive either hybrid closed-loop therapy or sensor-augmented pump therapy over 12 weeks of free living. Training on study insulin pump and continuous glucose monitoring took place over a 4-week run-in period. Eligible subjects were randomly assigned using central randomisation software. Allocation to the two study groups was unblinded, and randomisation was stratified within centre by low (<8·5%) or high (≥8·5%) HbA1c. The primary endpoint was the proportion of time that glucose concentration was within the target range of 3·9–10·0 mmol/L at 12 weeks post randomisation. Analyses of primary outcome and safety measures were done in all randomised patients. The trial is registered with ClinicalTrials.gov, number NCT02523131, and is closed to accrual.
From May 12, 2016, to Nov 17, 2017, 114 individuals were screened, and 86 eligible patients were randomly assigned to receive hybrid closed-loop therapy (n=46) or sensor-augmented pump therapy (n=40; control group). The proportion of time that glucose concentration was within the target range was significantly higher in the closed-loop group (65%, SD 8) compared with the control group (54%, SD 9; mean difference in change 10·8 percentage points, 95% CI 8·2 to 13·5; p<0·0001). In the closed-loop group, HbA1c was reduced from a screening value of 8·3% (SD 0·6) to 8·0% (SD 0·6) after the 4-week run-in, and to 7·4% (SD 0·6) after the 12-week intervention period. In the control group, the HbA1c values were 8·2% (SD 0·5) at screening, 7·8% (SD 0·6) after run-in, and 7·7% (SD 0·5) after intervention; reductions in HbA1c percentages were significantly greater in the closed-loop group compared with the control group (mean difference in change 0·36%, 95% CI 0·19 to 0·53; p<0·0001). The time spent with glucose concentrations below 3·9 mmol/L (mean difference in change −0·83 percentage points, −1·40 to −0·16; p=0·0013) and above 10·0 mmol/L (mean difference in change −10·3 percentage points, −13·2 to −7·5; p<0·0001) was shorter in the closed-loop group than the control group. The coefficient of variation of sensor-measured glucose was not different between interventions (mean difference in change −0·4%, 95% CI −1·4% to 0·7%; p=0·50). Similarly, total daily insulin dose was not different (mean difference in change 0·031 U/kg per day, 95% CI −0·005 to 0·067; p=0·09) and bodyweight did not differ (mean difference in change 0·68 kg, 95% CI −0·34 to 1·69; p=0·19). No severe hypoglycaemia occurred. One diabetic ketoacidosis occurred in the closed-loop group due to infusion set failure. Two participants in each study group had significant hyperglycaemia, and there were 13 other adverse events in the closed-loop group and three in the control group.
Hybrid closed-loop insulin delivery improves glucose control while reducing the risk of hypoglycaemia across a wide age range in patients with suboptimally controlled type 1 diabetes.
JDRF, NIHR, and Wellcome Trust.
Tight control of blood glucose concentration in people with type 1 diabetes predisposes to hypoglycaemia. We aimed to investigate whether day-and-night hybrid closed-loop insulin delivery can improve ...glucose control while alleviating the risk of hypoglycaemia in adults with HbA1c below 7·5% (58 mmol/mol).
In this open-label, randomised, crossover study, we recruited adults (aged ≥18 years) with type 1 diabetes and HbA1c below 7·5% from Addenbrooke's Hospital (Cambridge, UK) and Medical University of Graz (Graz, Austria). After a 2–4 week run-in period, participants were randomly assigned (1:1), using web-based randomly permuted blocks of four, to receive insulin via the day-and-night hybrid closed-loop system or usual pump therapy for 4 weeks, followed by a 2–4 week washout period and then the other intervention for 4 weeks. Treatment interventions were unsupervised and done under free-living conditions. During the closed-loop period, a model-predictive control algorithm directed insulin delivery, and prandial insulin delivery was calculated with a standard bolus wizard. The primary outcome was the proportion of time when sensor glucose concentration was in target range (3·9–10·0 mmol/L) over the 4 week study period. Analyses were by intention to treat. This study is registered with ClinicalTrials.gov, number NCT02727231, and is completed.
Between March 21 and June 24, 2016, we recruited 31 participants, of whom 29 were randomised. One participant withdrew during the first closed-loop period because of dissatisfaction with study devices and glucose control. The proportion of time when sensor glucose concentration was in target range was 10·5 percentage points higher (95% CI 7·6–13·4; p<0·0001) during closed-loop delivery compared with usual pump therapy (65·6% SD 8·1 when participants used usual pump therapy vs 76·2% 6·4 when they used closed-loop). Compared with usual pump therapy, closed-loop delivery also reduced the proportion of time spent in hypoglycaemia: the proportion of time with glucose concentration below 3·5 mmol/L was reduced by 65% (53–74, p<0·0001) and below 2·8 mmol/L by 76% (59–86, p<0·0001). No episodes of serious hypoglycaemia or other serious adverse events occurred.
Use of day-and-night hybrid closed-loop insulin delivery under unsupervised, free-living conditions for 4 weeks in adults with type 1 diabetes and HbA1c below 7·5% is safe and well tolerated, improves glucose control, and reduces hypoglycaemia burden. Larger and longer studies are warranted.
Swiss National Science Foundation (P1BEP3_165297), JDRF, UK National Institute for Health Research Cambridge Biomedical Research Centre, and Wellcome Strategic Award (100574/Z/12/Z).
Summary Background Closed-loop systems link continuous glucose measurements to insulin delivery. We aimed to establish whether closed-loop insulin delivery could control overnight blood glucose in ...young people. Methods We undertook three randomised crossover studies in 19 patients aged 5–18 years with type 1 diabetes of duration 6·4 years (SD 4·0). We compared standard continuous subcutaneous insulin infusion and closed-loop delivery (n=13; APCam01); closed-loop delivery after rapidly and slowly absorbed meals (n=7; APCam02); and closed-loop delivery and standard treatment after exercise (n=10; APCam03). Allocation was by computer-generated random code. Participants were masked to plasma and sensor glucose. In APCam01, investigators were masked to plasma glucose. During closed-loop nights, glucose measurements were fed every 15 min into a control algorithm calculating rate of insulin infusion, and a nurse adjusted the insulin pump. During control nights, patients' standard pump settings were applied. Primary outcomes were time for which plasma glucose concentration was 3·91–8·00 mmol/L or 3·90 mmol/L or lower. Analysis was per protocol. This trial is registered, number ISRCTN18155883. Findings 17 patients were studied for 33 closed-loop and 21 continuous infusion nights. Primary outcomes did not differ significantly between treatment groups in APCam01 (12 analysed; target range, median 52% IQR 43–83 closed loop vs 39% 15–51 standard treatment, p=0·06; ≤3·90 mmol/L, 1% 0–7 vs 2% 0–41, p=0·13), APCam02 (six analysed; target range, rapidly 53% 48–57 vs slowly absorbed meal 55% 37–64, p=0·97; ≤3·90 mmol/L, 0% 0–4 vs 0% 0–0, p=0·16), and APCam03 (nine analysed; target range 78% 60–92 closed loop vs 43% 25–65 control, p=0·0245, not significant at corrected level; ≤3·90 mmol/L, 10% 2–15 vs 6% 0–44, p=0·27). A secondary analysis of pooled data documented increased time in the target range (60% 51–88 vs 40% 18–61; p=0·0022) and reduced time for which glucose concentrations were 3·90 mmol/L or lower (2·1% (0·0–10·0) vs 4·1% (0·0–42·0); p=0·0304). No events with plasma glucose concentration lower than 3·0 mmol/L were recorded during closed-loop delivery, compared with nine events during standard treatment. Interpretation Closed-loop systems could reduce risk of nocturnal hypoglycaemia in children and adolescents with type 1 diabetes. Funding Juvenile Diabetes Research Foundation; European Foundation for Study of Diabetes; Medical Research Council Centre for Obesity and Related Metabolic Diseases; National Institute for Health Research Cambridge Biomedical Research Centre.
To evaluate feasibility, safety, and efficacy of day-and-night hybrid closed-loop insulin delivery in adolescents with type 1 diabetes under free-living conditions without remote monitoring or ...supervision.
In an open-label, randomized, free-living, crossover study design, 12 adolescents receiving insulin pump therapy (mean ±SD age 15.4 ± 2.6 years; HbA1c 8.3 ± 0.9%; duration of diabetes 8.2 ± 3.4 years) underwent two 7-day periods of sensor-augmented insulin pump therapy or hybrid closed-loop insulin delivery without supervision or remote monitoring. During the closed-loop insulin delivery, a model predictive algorithm automatically directed insulin delivery between meals and overnight; prandial boluses were administered by participants using a bolus calculator.
The proportion of time when the sensor glucose level was in the target range (3.9-10 mmol/L) was increased during closed-loop insulin delivery compared with sensor-augmented pump therapy (72 vs. 53%, P < 0.001; primary end point), the mean glucose concentration was lowered (8.7 vs. 10.1 mmol/L, P = 0.028), and the time spent above the target level was reduced (P = 0.005) without changing the total daily insulin amount (P = 0.55). The time spent in the hypoglycemic range was low and comparable between interventions.
Unsupervised day-and-night hybrid closed-loop insulin delivery at home is feasible and safe in young people with type 1 diabetes. Compared with sensor-augmented insulin pump therapy, closed-loop insulin delivery may improve glucose control without increasing the risk of hypoglycemia in adolescents with suboptimally controlled type 1 diabetes.
Human milk is the subject of many studies, but procedures for representative sample collection have not been established. Our improved methods for milk micronutrient analysis now enable systematic ...study of factors that affect its concentrations.
We evaluated the effects of sample collection protocols, variations in circadian rhythms, subject variability, and acute maternal micronutrient supplementation on milk vitamin concentrations.
In the BMQ (Breast-Milk-Quality) study, we recruited 18 healthy women (aged 18-26 y) in Dhaka, Bangladesh, at 2-4 mo of lactation for a 3-d supplementation study. On day 1, no supplements were given; on days 2 and 3, participants consumed ∼1 time and 2 times, respectively, the US-Canadian Recommended Dietary Allowances for vitamins at breakfast (0800-0859). Milk was collected during every feeding from the same breast over 24 h. Milk expressed in the first 2 min (aliquot I) was collected separately from the remainder (aliquot II); a third aliquot (aliquot III) was saved by combining aliquots I and II. Thiamin, riboflavin, niacin, and vitamins B-6, B-12, A, and E and fat were measured in each sample.
Significant but small differences (14-18%) between aliquots were found for all vitamins except for vitamins B-6 and B-12. Circadian variance was significant except for fat-adjusted vitamins A and E, with a higher contribution to total variance with supplementation. Between-subject variability accounted for most of the total variance. Afternoon and evening samples best reflected daily vitamin concentrations for all study days. Acute supplementation effects were found for thiamin, riboflavin, and vitamins B-6 and A at 2-4 h postdosing, with 0.1-6.17% passing into milk. Supplementation was reflected in fasting, 24-h postdose samples for riboflavin and vitamin B-6. Maximum amounts of dose-responding vitamins in 1 feeding ranged from 4.7% to 21.8% (day 2) and 8.2% to 35.0% (day 3) of Adequate Intake.
In the milk of Bangladeshi mothers, differences in vitamin concentrations between aliquots within feedings and by circadian variance were significant but small. Afternoon and evening collection provided the most-representative samples. Supplementation acutely affects some breast-milk micronutrient concentrations. This trial was registered at clinicaltrials.gov as NCT02756026.
Improvements in cardiovascular disease (CVD) rates among young adults in the past 2 decades have been offset by increasing racial/ethnic and gender disparities, persistence of unhealthy lifestyle ...habits, overweight and obesity, and other CVD risk factors. To enhance the promotion of cardiovascular health among young adults 18 to 39 years old, the medical and broader public health community must understand the biological, interpersonal, and behavioral features of this life stage. Therefore, the National Heart, Lung, and Blood Institute, with support from the Office of Behavioral and Social Science Research, convened a 2-day workshop in Bethesda, Maryland, in September 2017 to identify research challenges and opportunities related to the cardiovascular health of young adults. The current generation of young adults live in an environment undergoing substantial economic, social, and technological transformations, differentiating them from prior research cohorts of young adults. Although the accumulation of clinical and behavioral risk factors for CVD begins early in life, and research suggests early risk is an important determinant of future events, few trials have studied prevention and treatment of CVD in participants <40 years old. Building an evidence base for CVD prevention in this population will require the engagement of young adults, who are often disconnected from the healthcare system and may not prioritize long-term health. These changes demand a repositioning of existing evidence-based treatments to accommodate new sociotechnical contexts. In this article, the authors review the recent literature and current research opportunities to advance the cardiovascular health of today's young adults.