. The Monte Carlo simulation software is a valuable tool in radiation therapy, in particular to achieve the needed accuracy in the dose evaluation for the treatment plans optimisation. The current ...challenge in this field is the time reduction to open the way to many clinical applications for which the computational time is an issue. In this manuscript we present an innovative GPU-accelerated Monte Carlo software for dose valuation in electron and photon based radiotherapy, developed as an update of the FRED (Fast paRticle thErapy Dose evaluator) software.
. The code transports particles through a 3D voxel grid, while scoring their energy deposition along their trajectory. The models of electromagnetic interactions in the energy region between 1 MeV-1 GeV available in literature have been implemented to efficiently run on GPUs, allowing to combine a fast tracking while keeping high accuracy in dose assessment. The FRED software has been bench-marked against state-of-art full MC (FLUKA, GEANT4) in the realm of two different radiotherapy applications: Intra-Operative Radio Therapy and Very High Electron Energy radiotherapy applications.
. The single pencil beam dose-depth profiles in water as well as the dose map computed on non-homogeneous phantom agree with full-MCs at 2% level, observing a gain in processing time from 200 to 5000.
. Such performance allows for computing a plan with electron beams in few minutes with an accuracy of ∼%, demonstrating the FRED potential to be adopted for fast plan re-calculation in photon or electron radiotherapy applications.
Tumour control is performed in particle therapy using particles and ions, whose high irradiation precision enhances the effectiveness of the treatment, while sparing the healthy tissue surrounding ...the target volume. Dose range monitoring devices using photons and charged particles produced by the beam interacting with the patient's body have already been proposed, but no attempt has been made yet to exploit the detection of the abundant neutron component. Since neutrons can release a significant dose far away from the tumour region, precise measurements of their flux, production energy and angle distributions are eagerly sought in order to improve the treatment planning system (TPS) software. It will thus be possible to predict not only the normal tissue toxicity in the target region, but also the risk of late complications in the whole body. The aforementioned issues underline the importance of an experimental effort devoted to the precise characterisation of neutron production, aimed at the measurement of their abundance, emission point and production energy. The technical challenges posed by a neutron detector aimed at high detection efficiency and good backtracking precision are addressed within the MONDO (monitor for neutron dose in hadrontherapy) project, whose main goal is to develop a tracking detector that can target fast and ultrafast neutrons. A full reconstruction of two consecutive elastic scattering interactions undergone by the neutrons inside the detector material will be used to measure their energy and direction. The preliminary results of an MC simulation performed using the FLUKA software are presented here, together with the DSiPM (digital SiPM) readout implementation. New detector readout implementations specifically tailored to the MONDO tracker are also discussed, and the neutron detection efficiency attainable with the proposed neutron tracking strategy are reported.
The radiation used in hadrontherapy treatments interacts with the patient body producing secondary particles, either neutral or charged, that can be used for dose and Bragg peak monitoring and to ...provide a fast feedback on the treatment plans. Recent results obtained from the authors on simplified setups (mono-energetic primary beams interacting with homogeneous tissue-like target) have already indicated the correlation that exists between the flux of these secondaries coming from the target (e.g. protons and photons) and the position of the primary beam Bragg peak. In this paper, the measurements of charged particle fluxes produced by the interaction of a 220 MeV/u carbon ion beam at GSI, Darmstadt, with a polymethyl methacrylate target are reported. The emission region of protons (p), deuterons (d) and tritons (t) has been characterized using a drift chamber while the particle time-of-flight, used to compute the kinetic energy spectra, was measured with a LYSO scintillator. The energy released in the LYSO crystal was used for particle identification purposes. The measurements were repeated with the setup at 60° and 90° with respect to the primary beam direction. The accuracy on the fragments emission profile reconstruction and its relationship with the Bragg peak position have been studied. Based on the acquired experimental evidence, a method to monitor the dose profile and the position of the Bragg peak inside the target is proposed.
The high dose conformity and healthy tissue sparing achievable in Particle Therapy when using C ions calls for safety factors in treatment planning, to prevent the tumor under-dosage related to the ...possible occurrence of inter-fractional morphological changes during a treatment. This limitation could be overcome by a range monitor, still missing in clinical routine, capable of providing on-line feedback. The Dose Profiler (DP) is a detector developed within the INnovative Solution for In-beam Dosimetry in hadronthErapy (INSIDE) collaboration for the monitoring of carbon ion treatments at the CNAO facility (Centro Nazionale di Adroterapia Oncologica) exploiting the detection of charged secondary fragments that escape from the patient. The DP capability to detect inter-fractional changes is demonstrated by comparing the obtained fragment emission maps in different fractions of the treatments enrolled in the first ever clinical trial of such a monitoring system, performed at CNAO. The case of a CNAO patient that underwent a significant morphological change is presented in detail, focusing on the implications that can be drawn for the achievable inter-fractional monitoring DP sensitivity in real clinical conditions. The results have been cross-checked against a simulation study.
The MOnitor for Neutron Dose in hadrOntherapy (MONDO) project addresses the technical challenges posed by a neutron tracker detector aiming for a high detection efficiency and a good backtracking ...precision. The project aims to develop a tracking device capable of fully reconstructing the four momentum of fast and ultrafast secondary neutrons produced, e.g., in particle therapy (PT) treatments or in other physical processes. The MONDO tracker uses, as active material, squared scintillating fibers readout by dedicated CMOS-based digital single-photon avalanche diode (SPAD) array sensors. The expected light output, when operating in neutron monitoring applications, was experimentally evaluated in order to optimize the design of the MONDO detector readout. A small detector prototype (4 × 4 × 4.8 cm) has been built and tested at a test beam facility. The detection capabilities have been measured using a traditional photomultiplier (PMT) and a particle beam of 450-MeV electrons crossing a single layer of fibers. The observed number of photoelectrons in this case is (7.2 ± 1.4). A detector prototype was also tested with an SPAD-based SBAM (SPAD-Based Acquisition readout for MONDO experiment) sensor (SPADnet-I) to study the tracking performances. SBAM is a novel sensor developed to match the need of high single-photon detection efficiency and high spatial resolution and compactness. The sensor expected performance is discussed in view of an operation tailored for PT applications. In this contribution, we also report the results of a simulation performed to optimize the full MONDO detector layout.
The high dose conformity and healthy tissue sparing achievable in Particle Therapy when using C ions calls for safety factors in treatment planning, to prevent the tumor under-dosage related to the ...possible occurrence of inter-fractional morphological changes during a treatment. This limitation could be overcome by a range monitor, still missing in clinical routine, capable of providing on-line feedback. The Dose Profiler (DP) is a detector developed within the INnovative Solution for In-beam Dosimetry in hadronthErapy (INSIDE) collaboration for the monitoring of carbon ion treatments at the CNAO facility (Centro Nazionale di Adroterapia Oncologica) exploiting the detection of charged secondary fragments that escape from the patient. The DP capability to detect inter-fractional changes is demonstrated by comparing the obtained fragment emission maps in different fractions of the treatments enrolled in the first ever clinical trial of such a monitoring system, performed at CNAO. The case of a CNAO patient that underwent a significant morphological change is presented in detail, focusing on the implications that can be drawn for the achievable inter-fractional monitoring DP sensitivity in real clinical conditions. The results have been cross-checked against a simulation study.
The background induced by the high penetration power of the radiation is the main limiting factor of the current radio-guided surgery (RGS). To partially mitigate it, a RGS with β(+)-emitting ...radio-tracers has been suggested in literature. Here we propose the use of β(-)-emitting radio-tracers and β(-) probes and discuss the advantage of this method with respect to the previously explored ones: the electron low penetration power allows for simple and versatile probes and could extend RGS to tumours for which background originating from nearby healthy tissue makes probes less effective. We developed a β(-) probe prototype and studied its performances on phantoms. By means of a detailed simulation we have also extrapolated the results to estimate the performances in a realistic case of meningioma, pathology which is going to be our first in-vivo test case. A good sensitivity to residuals down to 0.1 ml can be reached within 1 s with an administered activity smaller than those for PET-scans thus making the radiation exposure to medical personnel negligible.