The KATRIN experiment is designed for a direct and model-independent determination of the effective electron anti-neutrino mass via a high-precision measurement of the tritium
β
-decay endpoint ...region with a sensitivity on
m
ν
of 0.2
eV
/
c
2
(90% CL). For this purpose, the
β
-electrons from a high-luminosity windowless gaseous tritium source traversing an electrostatic retarding spectrometer are counted to obtain an integral spectrum around the endpoint energy of 18.6 keV. A dominant systematic effect of the response of the experimental setup is the energy loss of
β
-electrons from elastic and inelastic scattering off tritium molecules within the source. We determined the energy-loss function in-situ with a pulsed angular-selective and monoenergetic photoelectron source at various tritium-source densities. The data was recorded in integral and differential modes; the latter was achieved by using a novel time-of-flight technique. We developed a semi-empirical parametrization for the energy-loss function for the scattering of 18.6-keV electrons from hydrogen isotopologs. This model was fit to measurement data with a 95%
T
2
gas mixture at 30 K, as used in the first KATRIN neutrino-mass analyses, as well as a
D
2
gas mixture of 96% purity used in KATRIN commissioning runs. The achieved precision on the energy-loss function has abated the corresponding uncertainty of
σ
(
m
ν
2
)
<
10
-
2
eV
2
1
in the KATRIN neutrino-mass measurement to a subdominant level.
The KArlsruhe TRItium Neutrino experiment (KATRIN) aims to determine the effective electron (anti)-neutrino mass with a sensitivity of 0.2eV/c
2
by precisely measuring the endpoint region of the ...tritium
β
-decay spectrum. It uses a tandem of electrostatic spectrometers working as magnetic adiabatic collimation combined with an electrostatic (MAC-E) filters. In the space between the pre-spectrometer and the main spectrometer, creating a Penning trap is unavoidable when the superconducting magnet between the two spectrometers, biased at their respective nominal potentials, is energized. The electrons accumulated in this trap can lead to discharges, which create additional background electrons and endanger the spectrometer and detector section downstream. To counteract this problem, “electron catchers” were installed in the beamline inside the magnet bore between the two spectrometers. These catchers can be moved across the magnetic-flux tube and intercept on a sub-ms time scale the stored electrons along their magnetron motion paths. In this paper, we report on the design and the successful commissioning of the electron catchers and present results on their efficiency in reducing the experimental background.
The KATRIN experiment aims to measure the effective electron antineutrino mass
m
ν
¯
e
with a sensitivity of
0.2
eV
/
c
2
using a gaseous tritium source combined with the MAC-E filter technique. A ...low background rate is crucial to achieving the proposed sensitivity, and dedicated measurements have been performed to study possible sources of background electrons. In this work, we test the hypothesis that gamma radiation from external radioactive sources significantly increases the rate of background events created in the main spectrometer (MS) and observed in the focal-plane detector. Using detailed simulations of the gamma flux in the experimental hall, combined with a series of experimental tests that artificially increased or decreased the local gamma flux to the MS, we set an upper limit of
0.006
count
/
s
(90% C.L.) from this mechanism. Our results indicate the effectiveness of the electrostatic and magnetic shielding used to block secondary electrons emitted from the inner surface of the MS.
The KATRIN experiment aims to determine the effective electron neutrino mass with a sensitivity of
0.2
eV/c
2
(%90 CL) by precision measurement of the shape of the tritium
β
-spectrum in the endpoint ...region. The energy analysis of the decay electrons is achieved by a MAC-E filter spectrometer. A common background source in this setup is the decay of short-lived isotopes, such as
219
Rn
and
220
Rn
, in the spectrometer volume. Active and passive countermeasures have been implemented and tested at the KATRIN main spectrometer. One of these is the magnetic pulse method, which employs the existing air coil system to reduce the magnetic guiding field in the spectrometer on a short timescale in order to remove low- and high-energy stored electrons. Here we describe the working principle of this method and present results from commissioning measurements at the main spectrometer. Simulations with the particle-tracking software
Kassiopeia
were carried out to gain a detailed understanding of the electron storage conditions and removal processes.
In this work, we present the first spectroscopic measurements of conversion electrons originating from the decay of metastable gaseous 83mKr with the Karlsruhe Tritium Neutrino (KATRIN) experiment. ...The obtained results represent one of the major commissioning milestones for the subsequent direct neutrino mass measurement with KATRIN. The successful campaign demonstrates the functionalities of the KATRIN beamline. Precise measurement of the narrow K-32, L3-32, and N2,3-32 conversion electron lines allowed to verify the eV-scale energy resolution of the KATRIN main spectrometer necessary for competitive measurement of the absolute neutrino mass scale.
The KATRIN experiment aims to determine the effective electron neutrino mass with a sensitivity of 0.2eV/c2 (90\% C.L.) by precision measurement of the shape of the tritium beta-spectrum in the ...endpoint region. The energy analysis of the decay electrons is achieved by a MAC-E filter spectrometer. A common background source in this setup is the decay of short-lived isotopes, such as 219Rn and 220Rn, in the spectrometer volume. Active and passive countermeasures have been implemented and tested at the KATRIN main spectrometer. One of these is the magnetic pulse method, which employs the existing air coil system to reduce the magnetic guiding field in the spectrometer on a short timescale in order to remove low- and high-energy stored electrons. Here we describe the working principle of this method and present results from commissioning measurements at the main spectrometer. Simulations with the particle-tracking software Kassiopeia were carried out to gain a detailed understanding of the electron storage conditions and removal processes.
The neutrino mass experiment KATRIN requires a stability of 3 ppm for the retarding potential at − 18.6 kV of the main spectrometer. To monitor the stability, two custom-made ultra-precise ...high-voltage dividers were developed and built in cooperation with the German national metrology institute Physikalisch-Technische Bundesanstalt (PTB). Until now, regular absolute calibration of the voltage dividers required bringing the equipment to the specialised metrology laboratory. Here we present a new method based on measuring the energy difference of two
83
m
Kr conversion electron lines with the KATRIN setup, which was demonstrated during KATRIN’s commissioning measurements in July 2017. The measured scale factor
M
=
1972.449
(
10
)
of the high-voltage divider K35 is in agreement with the last PTB calibration 4 years ago. This result demonstrates the utility of the calibration method, as well as the long-term stability of the voltage divider.