  • The HADES Pre-Shower detector
    Bałanda, A.; Jaskuła, M.; Kajetanowicz, M.; Kidoń, L.; Korcyl, K.; Kühn, W.; Kulessa, R.; Malarz, A.; Otwinowski, J.; Petri, M.; Pietraszko, J.; Prokopowicz, W.; Przygoda, W.; Salabura, P.; Skoczeń, A.; Szczybura, M.; Wajda, E.; Waluś, W.; Wiśniowski, M.; Wójcik, T.

    Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment, 10/2004, Letnik: 531, Številka: 3
    Journal Article

    The Pre-Shower detector was built for the high acceptance di-electron spectrometer to identify electrons produced in pion, proton and heavy ion-induced reactions in the 0.2<p<1.5 GeV/c momentum range. The system is composed of three layers of wire chambers interleaved with two layers of lead converters for electromagnetic shower formation. A scintillator wall is placed in front of the detector for time-of-flight measurements. The paper describes the Pre-Shower detector construction, the detector operation, the read-out electronics and its response to charged particles studied with electron, proton and heavy ion beams.